background image
Harvard Apparatus
phone
508.893.8999
toll free U.S.
800.272.2775
fax
508.429.5732
www.harvardapparatus.com
45
SYRINGE PUMP ACCESSORIES Pressure Transducers for PHD ULTRA
TM
CP

Syringe Pump
SYRINGE PUMP ACCESSORIES
Pressure Transducers for PHD ULTRA
TM
CP Syringe Pump
Order #
Product
PC5 73-0020
Blood Pressure Transducer P75 for PLUGSYS TAM Amplifier
PC5 73-3806
Blood Pressure Transducer P1500 for PLUGSYS TAM Amplifier
P75 Venous Pressure Transducer
& P1500 Pressure Transducer
P75 Venous Pressure Transducer
The P75 has a removable Macrolon
®
dome with a pressure connection
and a vent connection at the side, so that it can be filled free of air
bubbles. The dome connections have a male Luer taper so that suitable
stopcocks* can be attached. The transducer has a metal housing.
The actual pressure sensor inside is made from ceramic and therefore
has excellent resistance to different media. The transducer's rugged
construction can withstand pressure overloads up to 4000 mmHg
without damage. It works together with any DC bridge amplifier
(e.g., PLUGSYS TAM-A).
P1500 Pressure Transducer
For applications requiring constant pressure control in the 0-30
PSI (0-2 bar) range, we offer the P1500 Pressure Transducer. Like
the P75, it has a removable Macrolon
®
dome with a 90° pressure
connection and 45° vent connection on the top side for filling the
dome bubble free. The dome connections have a male Luer lock so
that suitable stopcocks* can be attached. The transducer has a metal
housing. The actual pressure sensor inside is made from ceramic and
therefore has excellent resistance to different media. The transducer
works with any DC bridge amplifier (e.g. PLUGSYS TAM-A) but is
supplied with the connector for the TAM. Configurations for other
amplifiers available upon request.
*UseoneeachofPC573-0096three-waystopcockandPC573-0097one-waystopcock
2F Pressure Catheter SPR-407
Millar Pressure Catheters
Order #
Model
Product
Adapters and Cables
PC5 72-9842
HEC-10C
Pressure Catheter with
Viking Connector
to PLUGSYS TAM Amplifier
PC5 72-9843
HEC-10D
Pressure Catheter with
Low Profile Connector
to PLUGSYS TAM Amplifier
Millar Pressure Catheters
PC5 72-9781
SPR-1000
1F, Single pressure, 20 cm, Straight,
Non-repairable
(Low Profile)
PC5 72-9779
SPR-671
1.4F Single pressure,1.1F-2F body
15 cm from tip, 65 cm
(Low Profile)
PC5 72-9775
SPR-407
2F Sensor, 1.5F Body, Single Pressure,
140 cm, Straight
(Viking)
PC5 72-9777
SPR-524
3.5F Single pressure, 100 cm, Straight,
Non-repairable
(Viking)
Millar pressure catheters continue to represent the gold standard
for high-fidelity pressure measurements for both systematic and
intraventricular applications.
Adapters and cables are available to connect catheters directly to the
range of Harvard Apparatus pressure amplifiers.
P75 and P1500 Pressure Transducers
Specifications
Model
P75 Pressure Transducer
P1500 Pressure Transducer
PRESSURE RANGE
±75 mmHg (±100 cmH
2
O)
±1,500 mmHg (±30 PSI)
OVERLOAD
-760 (=vacuum) to 4,000 mmHg
13,500 mmHg
SENSITIVITY
1 mV/mmHg, nominal
0.05 mV/mmHg, nominal
TEMPERATURE RANGE
0° to 50°C
-20° to 80°C
ZERO DRIFT
±0.04 mmHg/10°C (0° to 50°C)
±0.5% FS (-20° to 80°C)
RANGE DRIFT
±0.04 mV/10°C (±0.04 mmHg/10°C) (0° to 50°C)
±0.5% FS (-20° to 80°C)
LONG-TERM DRIFT
±0.04 mmHg FS/Year
±0.1% FS/Year
VOLUME DISPLACEMENT
0.06 mm
3
/10 mmHg
0.004 mm
3
/10 mmHg
FREQUENCY RANGE
0 to 100 Hz
0 to 300 Hz
OUTPUT RESISTANCE
300
, nominal
EXCITATION VOLTAGE
5 V (4.5 to 5.5 V) DC only
PRESSURE CONNECTIONS
Male Luer Lock
CONNECTION CABLE
1.5 m (4.9 ft) long
MOUNTING ROD, OD x L
8 x 70 mm (0.31 x 2.76 in)