background image
PHD 22/2000 MRI Compatible Syringe Pump
Harvard Apparatus
phone
508.893.8999
toll free U.S.
800.272.2775
fax
508.429.5732
www.harvardapparatus.com
SYRINGE PUMPS
PHD 22/2000 MRI Compatible Syringe Pump
34
The MRI Compatible Pump is based on the design of one of our most
tried and trusted syringe pumps. The PHD 22/2000. This pump is
made of mostly non-magnetic materials and can be placed near to an
imaging magnet when it is in use!
The syringe pumping mechanism
should be positioned at least 5 feet away from the opening of the
magnet and not in direct line of the opening.
The exact distance is
dependent upon the strength of the magnet. This will now make drug
or dye perfusions possible while the magnet is on. This pump is a
remote model where the syringe pumping mechanism is separated
from the control box by a 30 foot cable. This allows the pump to
be started and stopped near the magnet, via a RUN/STOP switch
that was added to the syringe holder. If additional space is required
between the control box and pumping mechanism, a 60 foot cable
is also available. For more details on the PHD 22/2000 Pump line,
see pages 32 to 35.
Program Description
The programming functions of this pump provide powerful capabilities
for advanced experiments. While in program mode this pump
can perform the following tasks at a predetermined time or when
prompted by a signal from an external device:
· Start or stop pumping
· Change pumping direction (infusion/withdrawal)
· Change flow rates
· Pump a precise volume and stop
· Pause operation
· Ramp up or down flow rates
In program mode the above tasks can be linked together into powerful
programs to simplify your automation projects.
Control Box
Syringe Holder
KEY FEATURES
High accuracy and precision
Can be placed near an imaging magnet
Holds 0.5 µl to 140 ml glass or plastic syringes
Legendary reliability ­ 2 year warranty
BENEFITS
Injecting Anesthesia
Drug Delivery
Injecting Dyes
PHD 22/2000
Specifications
TYPE
Microprocessor dual syringe infusion/withdrawal or
infusion/withdrawal programmable
ACCURACY
±0.35%
REPRODUCIBILITY
±0.05%
SYRINGE:
Type
Plastic or glass
Size Minimum
0.5 µl
Size Maximum
140 ml
FLOW RATE:
Minimum
0.0001 µl/hr
Maximum
220.82 ml/min
NON VOLATILE MEMORY
Stores all settings
CONNECTORS:
RS-232
RJ11-4 conductor
TTL
9-pin D-Sub Connector
AVERAGE LINEAR FORCE
22.7 kg (50 lbs)
STEP RESOLUTION
0.082 µm/µstep
VOLTAGE RANGE
Universal input 100/240 VAC, 50/60 Hz
CABLE LENGTH
9.1 m (30 ft)
DIMENSIONS, H x W x D:
Control Box
9.5 x 27.9 x 22.9 cm (3.75 x 11 x 9 in)
Syringe Holder
22.9 x 43.2 x 30.5 cm (9 x 17 x 12 in)
Order #
Product
PC5 70-2130
PHD 22/2000 MRI Compatible Remote Infusion/Withdrawal
Dual Syringe
PC5 70-2131
PHD 22/2000 MRI Compatible Remote Infusion/Withdrawal
Programmable Dual Syringe
SYRINGE PUMPS