background image
PHD ULTRA
TM
CP Constant Pressure Syringe Pump
Harvard Apparatus
phone
508.893.8999
toll free U.S.
800.272.2775
fax
508.429.5732
www.harvardapparatus.com
18
Specifications
PHD ULTRA
TM
CP PHD ULTRA
TM
CP
4400
TYPE
Microprocessor,
multiple syringe,
infusion/withdrawal,
programmable
Microprocessor,
multiple syringe,
infusion/withdrawal,
programmable
FLOW ACCURACY
±0.25%
±0.35%
PRESSURE ACCURACY
< ±2%
< ±2%
SYRINGE SIZE, MIX/
MAX
0.5 µl to 140 ml
0.5 µl to 140 ml
# SYRINGES
2-10
1
FLOW RATE:
MINIMUM
1.50 pl/min
3.06 pl/min
MAXIMUM
216.0 ml/min
216.0 ml/min
CONNECTORS:
USB (PC-to-Pump)
Type B
Type B
RS-232
(PC-to-Pump)
9-pin D-Sub
9-pin D-Sub
RS-485*
(Pump-to-Pump)
IEEE-1394, 6 pos
IEEE-1394, 6 pos
I/O & TTL
15-pin D-Sub
15-pin D-Sub
AVERAGE
LINEAR FORCE
34 kg (75 lbs)
at 100%
91 kg (200 lbs)
at 100%
MAXIMUM PRESSURE
(with 100 ml Syringe)
50 psi
135 psi
DIMENSIONS
CONTROL BOX
30.5 x 21.6 x 17.8 cm
(12 x 8.5 x 7 in)
30.5 x 21.6 x 10.6 cm
(12 x 8.5 x 4.3 in)
DIMENSIONS
REMOTE BOX
N/A
N/A
WEIGHT
4.5 kg (10 lbs)
5.4 kg (12 lbs)
VOLTAGE RANGE
100-240 VAC;
50/60 Hz
100-240 VAC;
50/60 Hz
Order #
PC5 88-3015
PC5 88-3016
A screen shot of the
Harvard Apparatus
Constant Pressure
Monitoring Software*.
PC5 88-3016 PHD ULTRA
TM
CP 4400
Infusion/Withdrawal Programmable
*Note: PC laptop not included.
SYRINGE PUMPS
PHD

ULTRA
TM
CP Constant Pressure Syringe Pump
SYRINGE PUMPS
KEY FEATURES
Enables continuous pressure-controlled infusion
Use in constant flow or constant pressure mode
Enhances safety for sensitive infusion targets and
physiological experimentation
Compatible with a wide range of pressure transducers
Automatically adjusts flow rate to maintain constant pressure
Alpha/numeric keyboard without a PC
Icon operation
New color LCD touch screen
Up-front control knobs for ease of operation
Vertical or horizontal orientation
Adjustable linear force to 75 lbs across the entire flow range
PC Application Included (For dataloging and PC control
if required)
CE, ETL (UL, CSA), WEEE, EU RoHS & CB Scheme Approved
2 year warranty
APPLICATIONS
Short-term organ perfusion studies that require the
maintenance of physiologic conditions.
Introduction of chemical reactants in a controlled manner.
Ocular injections and perfusions.
Small animal whole-body perfusions.
Constant pressure microfluidic mixing.
Administration of genetic material into organs without
viral vectors.