background image
Harvard Apparatus
phone
508.893.8999
toll free U.S.
800.272.2775
fax
508.429.5732
www.harvardapparatus.com
110
PUMP REFERENCES
Syringe Pump Pressure and Flow Rate
Harvard Apparatus
phone
508.893.8999
toll free U.S.
800.272.2775
fax
508.429.5732
www.harvardapparatus.com
Minimum/Maximum Flow Rates By Pump and Syringe Size
Pump 11 Elite Nanomite & PHD ULTRA
TM
Nanomite
Flow Rates
Nominal Minimum/Maximum Flow Rates for Various Syringes.
(Actual Limits will vary depending on syringe manufacturer)
Syringe Size
Diameter *
Minimum
Maximum
0.5 µl
0.103 mm
3.66 pl/min
1.909 µl/min
1 µl
0.146 mm
7.32 pl/min
3.819 µl/min
2 µl
0.206 mm
14.70 pl/min
7.635 µl/min
5 µl
0.343 mm
40.74 pl/min
21.17 µl/min
10 µl
0.485 mm
81.48 pl/min
42.32 µl/min
25 µl
0.729 mm
184.1 pl/min
95.62 µl/min
50 µl
1.030 mm
367.6 pl/min
190.9 µl/min
100 µl
1.457 mm
735.6 pl/min
381.9 µl/min
250 µl
2.304 mm
1.839 nl/min
955.1 µl/min
500 µl
3.256 mm
3.677 nl/min
1.907 ml/min
1000 µl
4.608 mm
7.358 nl/min
3.820 ml/min
* Note: These figures have been rounded and therefore may not exactly match the
Syringe Diameter Chart on page 112.
Pico Plus Elite
Flow Rates
Nominal Minimum/Maximum Flow Rates for Various Syringes.
(Actual Limits will vary depending on syringe manufacturer)
Syringe Size
Diameter *
Minimum
Maximum
0.5 µl
0.103 mm
0.54 pl/min
596.5 nl/min
1 µl
0.1457 mm
1.14 pl/min
1.194 µl/min
2 µl
0.206 mm
2.28 pl/min
2.386 µl/min
5 µl
0.343 mm
6.36 pl/min
6.615 µl/min
10 µl
0.485 mm
12.72 pl/min
13.230 µl/min
25 µl
0.729 mm
28.74 pl/min
29.880 µl/min
50 µl
1.030 mm
57.42 pl/min
59.650 µl/min
100 µl
1.457 mm
114.9 pl/min
119.4 µl/min
250 µl
2.304 mm
287.4 pl/min
298.5 µl/min
500 µl
3.256 mm
574.0 pl/min
596.1 µl/min
1000 µl
4.608 mm
1.150 nl/min
1.194 ml/min
1 ml
4.699 mm
1.196 nl/min
1.241 ml/min
3 ml
8.585 mm
3.990 nl/min
4.144 ml/min
5 ml
11.989 mm
7.782 nl/min
8.082 ml/min
10 ml
14.430 mm
11.270 nl/min
11.700 ml/min
* Note: These figures have been rounded and therefore may not exactly match the
Syringe Diameter Chart on page 112.
NanoCool Injector
Flow Rates
Nominal Minimum/Maximum Flow Rates for Various Syringes.
(Actual Limits will vary depending on syringe manufacturer)
Syringe Size
Diameter *
Minimum
Maximum
0.5 µl
0.103 mm
3.66 pl/min
1.909 µl/min
1 µl
0.146 mm
7.32 pl/min
3.819 µl/min
2 µl
0.206 mm
14.70 pl/min
7.635 µl/min
5 µl
0.343 mm
40.74 pl/min
21.17 µl/min
10 µl
0.485 mm
81.48 pl/min
42.32 µl/min
25 µl
0.729 mm
184.1 pl/min
95.62 µl/min
50 µl
1.030 mm
367.6 pl/min
190.9 µl/min
100 µl
1.457 mm
735.6 pl/min
381.9 µl/min
250 µl
2.304 mm
1.839 nl/min
955.1 µl/min
500 µl
3.256 mm
3.677 nl/min
1.907 ml/min
* Note: These figures have been rounded and therefore may not exactly match the
Syringe Diameter Chart on page 112.
PHD ULTRA
TM
4400
Flow Rates
Nominal Minimum/Maximum Flow Rates for Various Syringes.
(Actual Limits will vary depending on syringe manufacturer)
Syringe Size
Diameter *
Minimum
Maximum
0.5 µl
0.103 mm
3.0600 pl/min
1.59133 µl/min
1 µl
0.1457 mm
6.1200 pl/min
3.18423 µl/min
2 µl
0.206 mm
12.240 pl/min
6.36532 µl/min
5 µl
0.343 mm
33.960 pl/min
17.6471 µl/min
10 µl
0.485 mm
67.920 pl/min
35.2833 µl/min
25 µl
0.729 mm
153.480 pl/min
79.7151µl/min
50 µl
1.030 mm
306.420 pl/min
159.133 µl/min
100 µl
1.457 mm
613.200 pl/min
318.423 µl/min
250 µl
2.304 mm
1.53348 nl/min
796.252 µl/min
500 µl
3.256 mm
3.06258 nl/min
1.59021 ml/min
1 ml
4.699 mm
6.37872 nl/min
3.31205 ml/min
2.5 ml
4.851 mm
6.79806 nl/min
3.52979 ml/min
3 ml
8.585 mm
21.915 nl/min
11.0552 ml/min
5 ml
11.989 mm
41.5232 nl/min
21.5601 ml/min
8 ml
9.525 mm
26.2093 nl/min
13.6087 ml/min
10 ml
14.427 mm
60.1280 nl/min
31.2204 ml/min
20 ml
19.050 mm
104.837 nl/min
54.4347 ml/min
30 ml
21.590 mm
134.658 nl/min
69.9183 ml/min
50 ml
26.594 mm
204.311 nl/min
106.085 ml/min
100 ml
34.900 mm
351.865 nl/min
182.699 ml/min
140 ml
37.950 mm
416.009 nl/min
216.005 ml/min
* Note: These figures have been rounded and therefore may not exactly match the
Syringe Diameter Chart on page 112.