PLUGSYS pH Measurement Module (pHMM)

PLUGSYS pH Measurement Module (pHMM)

Used to measure pH with pH glass electrodes, primarily in biological fluids such as perfusate for isolated perfused organs.

Item# Description U.S. List Price Quantity
73-0215 pH Measurement Module (pHMM)
Login for U.S. Pricing


The pH Measurement Module (pHMM) is used to measure pH with pH glass electrodes. The main application is continuously pH recording with the combination pH electrodes for measurement in biological fluids such as perfusate for isolated perfused organs.

The input circuit of the module includes an isolation amplifier (potential separation between sensing electrode and circuit ground of the PLUGSYS measuring system) to avoid measurement errors due to ground loops and leakage currents.

Specifications
Input Connector     
Isolated BNC connector
pH Range 0 to 14
Resolution 0.01 pH
Output 1 V per 1 pH on BNC connector on front panel (±10 V);
output voltage is also available on PLUGSYS bus
Slope Adjustment    
2 point adjustment using calibrated buffer solutions
Simulation Physiological measuring range for calibrating
recording output can be simulated using 2
corresponding push buttons; each simulated value
can be adjusted in pH range of 0 to 14
PLUGSYS Width 2 slot units
Connector DIN 41612, 96-pin VG
DOCUMENT DESCRIPTION
73-0215_PHMM_ENG.pdfPLUGSYS pH Measurement Module (pHMM) User's Manual, Version 1.0