harvardapparatus.com is a data controller for the purposes of the General Data Protection Regulation ("GDPR"). By proceeding to tick the consent box at right, you consent to harvardbioscience.com processing your personal data, under the GDPR and any other applicable legislation, that we collect from you in accordance with our Privacy Policy. If you wish to withdraw your consent for us to process your personal data please contact our Compliance Team. Please address any questions, comments and requests (including your right to withdraw your consent to process your personal data) regarding our data processing practices to our Compliance Team.
Allow

PLUGSYS R-Wave Trigger Module (RWT)

Price on Request
SKU
1d-2f7000

Triggers (starts) a stimulator or any other device with the R-wave of an ECG signal. Requires ECGA module.

Grouped product items
Item No. Description Qty
73-1753 R-Wave Trigger Module (RWT) for ECG Amplifier Module (ECGA) Please login to add to Quote

The R-Wave Trigger Module (RWT) is especially designed to trigger (start) a stimulator or any other device with the R-wave of an ECG signal.

The ECG signal is measured with the PLUGSYS ECG Amplifier Module (ECGA) . The R-wave trigger module is internally connected to the ECG Amplifier and works only together with this amplifier.

The R-wave trigger module can be used (e.g., for synchronized stimulation of the heart) by triggering a delay module of a stimulator. It is possible to set the stimulation pulses delayed from the R-wave into the vulnerable phase to produce ventricular fibrillations. The trigger output of this module must be connected to the trigger input of a stimulator with a BNC cable.

Specifications73-1753
AccessoriesBNC output cable, operating instructions
Adjustable Blocking TimeBlocking time for trigger signal can be adjusted so that circuit is inactive for fixed time; time can be set with potentiometer 'BLOCKING TIME' in range from 50 msec to 500 msec; LED indicates blocking time
Dimensions (H x W x D)5.1 x 0.8 x 3.9 in (128.7 x 20.2 x 100 mm)
OutputBNC connector TRIGGER OUT at front panel open collector TTL signal, pull-up resistor 2.2 k; jumper on board defines whether positive or negative pulses used as trigger pulses
PLUGSYS Width*1 slot unit
Power SupplyThrough ECG Amplifier Type 689
R-Wave IndicationLED and beep to indicate R-wave; loudness of beep can be set with potentiometer
ThresholdThreshold for triggering can be set with potentiometer
Weight0.4 lb (0.2 kg)
AccessoriesBNC output cable, operating instructions