MouseOx Plus Pulse Oximeter for Rodents

MouseOx Plus Complete System

Non-invasive pulse oximeter providing real–time, continuous measurements of arterial O2 saturation, heart rate, breath rate, temperature, pulse distention, breath distention and activity in conscious and anesthetized rodents, from neonatal mice through adult rats

Item# Description U.S. List Price Quantity
72-8010 MouseOx®Plus Pulse Oximeter, 115 VAC, 60 Hz (Basic System)
72-8011 MouseOx®Plus Pulse Oximeter, 230 VAC, 50 Hz (Basic System)
72-8018 MouseOx®Plus Pulse Oximeter for Rodents with Software, 115 VAC, 60 Hz (Premium Bundle)
72-8019 MouseOx®Plus Pulse Oximeter for Rodents with Software, 230 VAC, 50 Hz (Premium Bundle)
72-8020 MouseOx®Plus Pulse Oximeter with Software and STARR-Link Analog Output Module, 115 VAC, 60 Hz (Premium Bundle)
72-8021 MouseOx®Plus Pulse Oximeter with Software and STARR-Link Analog Output Module, 230 VAC, 50 Hz (Premium Bundle)
72-8013 16-Channel Multiplexer for MouseOxPlus, 115 VAC, 60 Hz
72-8014 16-Channel Multiplexer for MouseOxPlus, 230 VAC, 50 Hz
72-8083 8-Channel Multiplexer for MouseOxPlus, 115 VAC, 60 Hz
72-8084 8-Channel Multiplexer for MouseOxPlus, 230 VAC, 50 Hz
72-8008 8-Channel Multiplexer for MouseOxPlus, with Temperature Measurement, 115 VAC, 60 Hz
72-8009 8-Channel Multiplexer for MouseOxPlus, with Temperature Measurement, 230 VAC, 50 Hz
72-8080 Universal Cable, 1.8 m (6 ft)
72-8081 Universal Cable, 3.7 m (12 ft)
75-0550 MouseOx®Plus Mouse Conscious Monitoring Equipment, to mount onto 61-0046
72-8012 MouseOx®Plus Premium Application Software
72-8016 MouseOx®Plus, Temperature Sensor & Software Upgrade Module, with 6 ft (180 cm) Cable
72-9709 MouseOx®Plus Conscious Monitoring Module
72-8017 MouseOx®Plus, MRI Temperature Sensor & Software Upgrade Module, with 20 ft (6.1 m) MRI Cable
72-8022 STARR-Link™ Analog Data Output Module for MouseOx®Plus, 115 VAC, 60 Hz
72-8023 STARR-Link™ Analog Data Output Module for MouseOx®Plus, 230 VAC, 50 Hz
Login for U.S. Pricing


This non-invasive pulse oximeter provides real–time, continuous physiological measurements. It is designed to work with neonatal mice through adult rats. It has a patented design to detect heart rates of 90-900 bpm. The MouseOxPlus® easily allows researchers to titrate mechanical ventilation and monitor depth of anesthesia.

A complete system for anesthetized animals consists of the MouseOx Plus, software, and a sensor. A complete System for conscious animals consists of the MouseOx Plus, software with Conscious Monitoring module, conscious monitoring equipment and a collar clip sensor.

  • Option for Premium software or DAQ interface boxfor data collection and analysis
  • Option for multiple animal monitoring from one MouseOx Plus system

All MouseOx Plus® systems come with two sensors of your choice.

Provides the following measurements:

  • Arterial Oxygen Saturation
  • Heart Rate– Breath Rate
  • Temperature (optional)
  • Pulse Distention
  •  Breath Distention
  • Activity (optional)

MouseOx and MouseOxPlus are registered trademarks of Starr Life Sciences. ThroatClip, STARR-link and CollarClip are trademarks of Starr Life Sciences.

Specifications
Oxygen Saturation (SpO2):  
      Range 0 to 100% Arterial Blood Oxygen Saturation 
      Resolution <1.5% across entire measurement range
      Response Time Reported after every heartbeat; screen refreshed every 0.72 seconds 
 Pulse Rate (BPM):
      Range
 90 to 900 BPM
  Pulse Rate (BPM)  Resolution (BPM)
   210 2.4 
   300 4.9
   400  8.7
   500 13.5
  600   19.4
   700  26.2
   800  34
   900  42.9
      Response Time
 Reported after every heartbeat; screen refreshed every 0.72 seconds
 Breath Rate (BrPM):  
      Range  25 to 450 BrPM  
   Breath Rate (BrPM)  Resolution (BrPM)
   25  0.1
   100  0.6
   200  2.2
   300  4.9
   400  8.7
      Response Time  Reported every 1.7 seconds as a moving average of the rate of breathing for the previous 10 breath rate measurements.The screen updates every 0.72 seconds, butnew breath rate data is presented at the lower rate.
 Pulse Distention:
      Range
 0 to 800 mm
 
      Resolution
  <2.4% of measurement
 
      Response Time
  Reported after every heartbeat; screen refreshed every 0.72 seconds

 Breath Distention:
   
      Range
  0 to 800 mm
 
      Resolution 
  <2.4% of measurement
 
      Response Time
  Reported every 1.7 seconds
 
 Temperature Range
 15° to 50°C 
 Resolution
  ±0.1 °C
 Response Time
  Reported every 0.72 seconds