harvardapparatus.com is a data controller for the purposes of the General Data Protection Regulation ("GDPR"). By proceeding to tick the consent box at right, you consent to harvardbioscience.com processing your personal data, under the GDPR and any other applicable legislation, that we collect from you in accordance with our Privacy Policy. If you wish to withdraw your consent for us to process your personal data please contact our Compliance Team. Please address any questions, comments and requests (including your right to withdraw your consent to process your personal data) regarding our data processing practices to our Compliance Team.
Allow

PLUGSYS Ventilation Sequencer Module (VSM)

Price on Request
SKU
d36f9bdf
For applications for which multiple rodents must be ventilated simultaneously (e.g. multiple blood pressure measurement) or Einthoven bronchospasmolysis test on multiple animals
Grouped product items
Item No. Description Qty
No options of this product are available.

The Ventilation Sequencer Module (VSM) consists of an electronic and a pneumatic section and requires a compressed air supply. It is used to set the respiratory rate, inspiratory:expiratory (I:E) ratio and the maximum inspiratory pressure (i.e. safety pressure) for all animals being ventilated.

One to eight ventilation heads 73-0047 can be connected to the VSM, allowing ventilation of up to eight rodents at a time. Tidal volume can be adjusted individually.

Principle of Operation

The VSM module with the ventilation head operates on the principle of ‘intermittent constant flow ventilation’. A continuously adjustable air flow is passed by a main valve either to the animal (during inspiration) or to an overflow outlet (during expiration). The design of the main valve provides an adjustable, pressure controlled, protection against excessive pressure rise in the animal’s lung. The main valve is controlled electronically according to the selected respiration rate (RATE) and the selected inspiratory cycle (INSP. CYCLE %). The safety pressure which is the maximum admitted inspiratory pressure can be adjusted on the VSM module. The timing and the safety pressure are the same for all the ventilation heads, the constant air flow during inspiration can be adjusted individually on each ventilation head.

The safety pressure is set by the researcher and is used to prevent over inflation of the animal’s lungs. The VSM requires a compressed air supply and, using both electronic and pneumatic controls, directs the air flow to the VSM ventilation head.  If more than one ventilation head is used, a connecting block 73-0048 is required.

A directional valve within the ventilation head is controlled by the VSM to regulate the inspiratory and expiratory cycle. The VSM ventilation head includes all necessary tubing.

An optional MPX pressure transducer, 73-0064, can be connected to the ventilation head. This pressure transducer attached to a TAM-A transducer amplifier, 73-0065, can be used to record tracheal/airway pressures. This setup is described more thoroughly in the Einthoven Anti-asthmatics system.  All components must be purchased separately.

Compressed Air Supply:
Minimum 4 bar (58 lb)
Maximum 8 bar (116 lb)
Respiratory Rate 30, 40, 50, 60, 70, 80, 90, 100 breaths/min
Inspiratory Time 10, 20, 30, 40, 50, 60, 70, 80 % of total cycle duration
Control of Ventilation Heads:
Ventilation Pressure 150 mmHg
Control Pressure Adjustable 0 to 50 cmH2O, safety pressure set at Max. Insp. Pressure
Ventilator Sequencer Module (VSM)
Indication LED lights up during inspiration
Trigger Output BNC socket on front panel (TTL open collector; inspiration = low level; expiration = high level); trigger signal also available on PLUGSYS bus system
Dimensions, H x W x D 1128.7 x 101.6 x 220 mm (5.1 x 4 x 8.7 in)
PLUGSYS Width 5 slot units
Connector DIN 41612, 96-pin VG