array(3) { ["show"]=> bool(true) ["ip"]=> string(13) "62.218.203.35" ["country"]=> string(2) "AT" } desde la funcion de ip obtenemos IP invalida
harvardapparatus.com is a data controller for the purposes of the General Data Protection Regulation ("GDPR"). By proceeding to tick the consent box at right, you consent to harvardbioscience.com processing your personal data, under the GDPR and any other applicable legislation, that we collect from you in accordance with our Privacy Policy. If you wish to withdraw your consent for us to process your personal data please contact our Compliance Team. Please address any questions, comments and requests (including your right to withdraw your consent to process your personal data) regarding our data processing practices to our Compliance Team.
Allow

PLUGSYS Monophasic Action Potential Module (MAPM)

Price on Request
SKU
GRP-MAPM

Use for capturing and amplifying monophasic action potentials (MAPs)

Grouped product items
Item No. Description Qty
73-1780 Monophasic Action Potential Module (MAPM), 6-Channel Please login to add to Quote
73-1787 Input Box for up to 6 MAP Channels, for 1 PLUGSYS MAPM Module, with Connection Cables Please login to add to Quote

The Monophasic Action Potential Module (MAP) is for capturing and amplifying monophasic action potentials (MAPs) in the PLUGSYS system. It has to be connected to special MAP electrodes. This module incorporates isolated input amplifiers in order to avoid hum interference. In this way the inputs are isolated electrically from the output circuits and from the housing.

The output voltages of the amplified MAP signals can be switched with jumpers to the PLUGSYS system bus. From the system bus they can be picked off by a Recorder Output Module (ROM) for direct recording or a DAQ Hardware PLUGSYS for data acquisition by computer. A square-wave calibration generator with a 10 mV amplitude is incorporated for calibrating the amplitudes.

Inputs: A special connection box is required for connecting the MAP electrodes, with separate switch-off facility for the individual channels. The connection box has to be placed close to the electrodes so that the unscreened electrode leads can be kept as short as possible. The box is not part of the module and has to be ordered separately.

Specifications73-178073-3080
Amplifier TypeDC amplifier with automatic zeroing, 6 channelsDC amplifier with automatic zeroing, 6 channels
CalibrationSquare-wave signal 10 mV 2 HzSquare-wave signal 10 mV 2 Hz
Commond Mode RejectionBetter than 1 in 50,000 at 50 HzBetter than 1 in 50,000 at 50 Hz
Connector96-pin VG connector to DIN 41612 submin-D 15-HD96-pin VG connector to DIN 41612 submin-D 15-HD
FilterLow-pass filter: 1 kHz, 2 kHz and 5 kHzLow-pass filter: 1 kHz, 2 kHz and 5 kHz
GainSetting x0.5 - 100; Setting x1 - 200; Setting x2 - 400Setting x0.5 - 100; Setting x1 - 200; Setting x2 - 400
Input Box(Order separately) input box with 6 differential inputs, sockets suitable for 2 mm dia. plug pins, including approx. 2 m connecting cable(Order separately) input box with 6 differential inputs, sockets suitable for 2 mm dia. plug pins, including approx. 2 m connecting cable
Input CircuitDifferential inputs, isolated (floating input), isolating voltage 200 V DC max.Differential inputs, isolated (floating input), isolating voltage 200 V DC max.
Input ImpedanceWhen switched on (INPUT ON): 1010Ω
When switched off (INPUT OFF): 5000Ω approx.
When switched on (INPUT ON): 1010Ω
When switched off (INPUT OFF): 5000Ω approx.
OutputOutput signals of 6 MAP signals are linked by jumpers to PLUGSYS bus system; link to display unit is provided by Recorder Output Module (ROM) installed in PLUGSYS systemOutput signals of 6 MAP signals are linked by jumpers to PLUGSYS bus system; link to display unit is provided by Recorder Output Module (ROM) installed in PLUGSYS system
PLUGSYS Width*3 slot units3 slot units
Power Supply5 V 1.2 mA from PLUGSYS system bus5 V 1.2 mA from PLUGSYS system bus
ZeroingVoltage range Voltage range