PLUGSYS Monophasic Action Potential Module (MAPM)
Use for capturing and amplifying monophasic action potentials (MAPs)
The Monophasic Action Potential Module (MAP) is for capturing and amplifying monophasic action potentials (MAPs) in the PLUGSYS system. It has to be connected to special MAP electrodes. This module incorporates isolated input amplifiers in order to avoid hum interference. In this way the inputs are isolated electrically from the output circuits and from the housing.
The output voltages of the amplified MAP signals can be switched with jumpers to the PLUGSYS system bus. From the system bus they can be picked off by a Recorder Output Module (ROM) for direct recording or a DAQ Hardware PLUGSYS for data acquisition by computer. A square-wave calibration generator with a 10 mV amplitude is incorporated for calibrating the amplitudes.
Inputs: A special connection box is required for connecting the MAP electrodes, with separate switch-off facility for the individual channels. The connection box has to be placed close to the electrodes so that the unscreened electrode leads can be kept as short as possible. The box is not part of the module and has to be ordered separately.
Specifications | 73-1780 | 73-3080 |
---|---|---|
Amplifier Type | DC amplifier with automatic zeroing, 6 channels | DC amplifier with automatic zeroing, 6 channels |
Calibration | Square-wave signal 10 mV 2 Hz | Square-wave signal 10 mV 2 Hz |
Commond Mode Rejection | Better than 1 in 50,000 at 50 Hz | Better than 1 in 50,000 at 50 Hz |
Connector | 96-pin VG connector to DIN 41612 submin-D 15-HD | 96-pin VG connector to DIN 41612 submin-D 15-HD |
Filter | Low-pass filter: 1 kHz, 2 kHz and 5 kHz | Low-pass filter: 1 kHz, 2 kHz and 5 kHz |
Gain | Setting x0.5 - 100; Setting x1 - 200; Setting x2 - 400 | Setting x0.5 - 100; Setting x1 - 200; Setting x2 - 400 |
Input Box | (Order separately) input box with 6 differential inputs, sockets suitable for 2 mm dia. plug pins, including approx. 2 m connecting cable | (Order separately) input box with 6 differential inputs, sockets suitable for 2 mm dia. plug pins, including approx. 2 m connecting cable |
Input Circuit | Differential inputs, isolated (floating input), isolating voltage 200 V DC max. | Differential inputs, isolated (floating input), isolating voltage 200 V DC max. |
Input Impedance | When switched on (INPUT ON): 1010Ω
When switched off (INPUT OFF): 5000Ω approx. | When switched on (INPUT ON): 1010Ω
When switched off (INPUT OFF): 5000Ω approx. |
Output | Output signals of 6 MAP signals are linked by jumpers to PLUGSYS bus system; link to display unit is provided by Recorder Output Module (ROM) installed in PLUGSYS system | Output signals of 6 MAP signals are linked by jumpers to PLUGSYS bus system; link to display unit is provided by Recorder Output Module (ROM) installed in PLUGSYS system |
PLUGSYS Width* | 3 slot units | 3 slot units |
Power Supply | 5 V 1.2 mA from PLUGSYS system bus | 5 V 1.2 mA from PLUGSYS system bus |
Zeroing | Voltage range | Voltage range |