Flow Measurement Options for UP-100IH, IH-SR and IH-5 in Langendorff and Working Heart Modes
Flow measurement can be accomplished in a functional IH system using two methods:
- Indirect measurement by controlling the pump speed of a roller pump (SCP controller)
- Direct measurement with ultrasound transit time technique (TTFM-2 module)
Required components are included in the IH-SR or IH-5 Core Systems and the selected IH-SR or IH-5 Core Systems Options. Species-specific flow probes and adapters are also required.
Flow measurement can be accomplished in a functional IH system using two methods:
- Indirect measurement by controlling the pump speed of a roller pump (SCP controller)
- Direct measurement with Ultrasound Transit Time Technic (TTFM-2 module)
Applications
- Accurate coronary flow measurement on drug studies
- Studies on coronary resistance changes (pre and post ischemia)
- Myogene autoregulation (reactive hyperemia)
Constant Pressure/Flow Control
If you do "Langendorff only", the PLUGSYS Servo Controller Module (SCP) is used in combination with a roller pump to perfuse the heart in constant pressure mode. The SCP module calculates the coronary flow from the pump speed of the roller pump (indirect flow measurement). The PLUGSYS TAM-D module amplifies the measured perfusion pressure signal and outputs it to the data acquisition system. Both the roller pump and the TAM-D are included in the IH-SR and IH-5 Core Systems.
- Allows perfusion under constant pressure or constant flow (direct switchable)
- Provides accurate control of perfusion pressure or flow, even at very low flow rates
- Indirect coronary flow measurement from pump speed
Direct Flow Measurement with Ultrasound Transit Time Flowmeter
The PLUGSYS TTFM-2 module, in combination with a Starling Pressure Controller, provides real and accurate transit time flow measurement in ml/min for direct coronary flow measurement in retrograde Langendorff mode or aortic flow measurement in working ejecting heart. In addition, species-specific flow probes and adapters are required.
Features & Benefits
- Accurate ultrasonic transit time direct flow measurement
- Unique integration of flow probe into perfusion path
- Laminar flow lines improve accuracy of flow measurement
- Thermal properties of Perspex adapter reduce temperature loss
- Increased system flexibility by allowing the study of Myogene Autoregulation (Reactive Hyperemia)
- Indirect flow measurement can typically follow increasing flow rates but often lags behind decreasing flow rates
The table below specifies the flow probe and adapter to be used for each type of isolated heart system (IH-SR, IH-5 or UP-100IH) for the indicated application. It is extremely important to order the correct probe and adapter pair. If ordering online, please select carefully from the item listing or contact technical support for assistance.
Type of Measurement/Species | Required Probe and Adapter |
Direct Flow Measurement/Mouse with IH-SR Application: Coronary flow measurement in retrograde Langendorff mode or aortic flow measurement in working heart. |
73-4668 MA1PRB Flow Probe for TTFM-2 73-3554 Adapter for 73-4668 1RB Flow Probe for IH-SR (flow probe is embedded into this adapter) |
Direct Flow Measurement/Rat with IH-SR Application: Coronary flow measurement in retrograde Langendorff mode or aortic flow measurement in working heart. |
73-4644 MA2PSB Flow Probe for TTFM-2 73-2819 Adapter for 73-4644 2SB Flow Probe for IH-SR (flow probe is embedded into this adapter) |
Direct Flow Measurement/Rat and Guinea Pig with IH-5 Application: Coronary flow measurement in retrograde Langendorff mode or aortic flow measurement in working heart. |
73-4647 MA2.5PSB Flow Probe for TTFM-2 or TS420 Module |
Direct Flow Measurement/Rabbit with IH-5 Application: Coronary flow measurement in retrograde Langendorff mode or aortic flow measurement in working heart. |
73-4652 MA4PSB Flow Probe for TTFM-2 or TS420 Module 73-3069 Adapter for 73-4652 4SB Flow Probe for IH-5 |
Cardiac Output Measurement, Working Heart/Mouse with IH-SR Application: Flow measurement into left atrium in Working Heart mode |
73-4668 MA1PRB Flow Probe for TTFM-2 73-3554 Adapter for 73-4668 1RB Flow Probe for IH-SR (flow probe is embedded into this adapter) |
Cardiac Output Measurement, Working Heart/Rat with IH-SR Application: Flow measurement into left atrium in Working Heart mode |
73-4647 MA2.5PSB Flow Probe for TTFM-2 73-2820 Adapter for 73-4647 2.5SB Flow Probe for IH-SR (flow probe is embedded into this adapter) |
Cardiac Output Measurement, Working Heart/Rat and Guinea Pig with IH-5 Application: Flow measurement into left atrium in Working Heart mode |
73-4647 MA2.5PSB Flow Probe for TTFM-2 or TS420 module 73-3071 Adapter 73-4647 2.5SB Flow Probe 2.5SB for IH-5 |
Cardiac Output Measurement, Working Heart, Rabbit with IH-5 Application: Flow measurement into left atrium in Working Heart mode |
73-4652 MA4PSB Flow Probe for TTFM-2 or TS420 Module 73-3069 Adapter for 73-4652 4SB Flow Probe for IH-5 |