array(3) { ["show"]=> bool(true) ["ip"]=> string(13) "74.80.208.120" ["country"]=> string(2) "US" } desde la funcion de ip obtenemos IP invalida
harvardapparatus.com is a data controller for the purposes of the General Data Protection Regulation ("GDPR"). By proceeding to tick the consent box at right, you consent to harvardbioscience.com processing your personal data, under the GDPR and any other applicable legislation, that we collect from you in accordance with our Privacy Policy. If you wish to withdraw your consent for us to process your personal data please contact our Compliance Team. Please address any questions, comments and requests (including your right to withdraw your consent to process your personal data) regarding our data processing practices to our Compliance Team.
Allow

Standard Flow Probes (R-Series - J Reflector)

Price on Request
SKU
GRP-RPROBES

Limited available quantities.  Please contact us for further information.

For direct measurement of volume flow for vessels from 1 to 1.5 mm diameter using the PLUGSYS transit time flow meter module (TTFM-2). These probes have the transducers positioned at an 45° angle to the vessel. The probes are available with the cable exit in the back or on the side.

For information on flow probes to interface with the earlier versions of the TTFM (73-0146) please contact Technical Services.

Grouped product items
Item No. Description Qty
73-4669 MA1PRS Flow probe for HSE TTFM-2 or TS420 module Please login to add to Quote
73-4670 MA1.5 PRB Flow probe for TTFM-2 or TS420 Module Please login to add to Quote

For direct measurement of volume flow for vessels from 1 to 1.5 mm diameter using the PLUGSYS transit time flow meter module (TTFM-2). These probes have the transducers positioned at an 45° angle to the vessel. The probes are available with the cable exit in the back or on the side.

Coding: 

  • RB:  R-probe with cable exit in the Back
  • RS:  R-probe with cable exit on the Side 

All R Probes do have a J reflector with slide cover.

For information on flow probes to interface with the earlier versions of the TTFM (73-0146) please contact Technical Services.

Item Number 73-4668 73-4669 73-4670
Diameter 1 mm 1 mm 1.5 mm
OD Vessel Acute 0.7 - 1.2 0.7 - 1.2 1.2 - 1.8
Resolution ml/min 0.05 0.05 0.075

Range ml/min

Sc. Low

 

5

 

5

 

10

Sc. Norm  20 20 40
Max 100 100 200
Probe Length  6.5 mm 6.5 mm 7.6 mm
Probe Width 4 mm 4 mm 3.5 mm
Cable Length  60 cm  60 cm  60 cm 
Cable Exit  Back  Side  Back