Home Syringe Pumps/Peristaltic Pumps Syringe Pumps OEM Pump Modules High Pressure OEM Syringe Pump Modules


Larger Views

 

High Pressure OEM Syringe Pump Modules
The High Pressure Syringe Pump Module employs a microcontroller which controls a small step angle stepping motor that drives a lead screw and Pusher Block. Microstepping techniques are employed to further reduce the step angle, eliminating flow pulsation. Data can be entered via an RS-232 connector located on the micro controller. The microcontroller calculates the cross-sectional area of the syringe selected and calibrates the flow rate and volume accumulation. The numerous features of the High Pressure Pump Module result from the use of microprocessor technology.
Search for Product in current catalog
Item Listing Full Description Specifications Journal Articles Manuals
Item# Description In Stock Lead Time U.S. List Price Quantity
702202 High Pressure Syringe Pump Module -- 5 days View Pricing
Login to Shop

For local pricing, contact your nearest distributor.



Full Description Back to Top

• Easy to incorporate legendary syringe pump technology into your equipment
• Ideal high force syringe pump for do-it-yourselfers and OEM equipment designers
• Precisely dispenses volumes at high pressure
• Smooth flow
• Nonvolatile memory
• Three modes of operation:
- Constant flow rate
- Volume dispense
- Programmable
• Quantity discounts available on request

The High Pressure Syringe Pump Module employs a microcontroller which controls a small step angle stepping motor that drives a lead screw and Pusher Block. Microstepping techniques are employed to further reduce the step angle, eliminating flow pulsation. Data can be entered via an RS-232 connector located on the micro controller. The microcontroller calculates the cross-sectional area of the syringe selected and calibrates the flow rate and volume accumulation. The numerous features of the High Pressure Pump Module result from the use of microprocessor technology.

The High Pressure Programmable Pump Module model provides full programmability along with Infuse/Withdraw capability. This unit is designed to operate inside an enclosure, cabinet, or on top of a bench. The board may be removed for “remote” operation.

Pressure and Speed
The High Pressure Pump Module can deliver up to 220.82ml/minute with a single 140ml syringe. Maximum pressure is dependent on syringe size. Drive produces >200 lbs linear force.

Infusion and Refill Rates
Specify independent rates for infusing and refilling. This allows a slow infusion rate then a fast refill.

Target Volume
Specify the volume that is to be infused or refilled. The pump will run at the rate specified until this volume has been delivered when in the Volume mode.

Auto Fill
Auto Fill automatically activates an externally attached solenoid (refer to Appendix L for part number) and refills the syringe when it is empty. This permits infusions to be virtually independent of syringe capacity.

Modes of Operation
(Set for pump mode, can be changed thru RS-232)
Pump: Runs continuously in the infuse or refill directions until stopped.
Volume: Runs until a specified volume has been pumped or refilled.
Program: Pump operates according to specified sequence of instructions.
(Note: All modes interact with Auto Fill)

External Connections User I/O
Allows pump operations to be synchronized with external devices or by a person at a distance from the pump. Connector pins are available to control direction of pump travel to control an external valve for refilling, and for general use. A simple contact closure to ground or TTL level signals may be used for inputs.

RS-232
Multiple pumps can be ‘daisy chained’ together and remotely controlled from a computer or any device communicating via RS-232.

A scale can be connected, enabling the pump to infuse by weight instead of by volume.
A printer can be connected to record final volumes or weights whenever the pump stops. In addition the program entered for the program mode can be listed on a connected printer. Both a scale and a printer may be connected simultaneously.

Non-Volatile Memory
All operational data entered into the pump from a computer will be stored, including the program.

Stall Detection
An optical detector is used to verify expected movement of the motor. If the motor is prevented from turning due to jamming or excessive back pressure, the pump will stop.

Program Storage
Programmable model can store up to 4 sets of 9 program sequences for later selection.

Infuse Rate
The Infuse Rate is the rate of pumping while infusing in the Pump or Volume modes.

Target Volume
The Target Volume is the volume that you desire to deliver at the set Infuse or Refill Rate.

Auto Fill
When set to ‘ON’, the syringe is assumed to be empty. Auto Fill continuously monitors the volume of the syringe according to the volume pumped. When the pump determines that the syringe is empty, the operation in progress is suspended and Auto Fill is activated. The pumping direction is then reversed and the pump runs at the refill rate.

Program Description
The programming functions of this pump provide powerful capabilities for advanced experiments. While in program mode this pump can perform the following tasks at a predetermined time or when prompted by a signal from an external device:
- start or stop pumping
- change pumping direction (infuse-withdraw)
- change flow rates
- pump a precise volume and stop
- pause operation
- ramp up or down flow rates

In program mode the above tasks can be linked together into powerful programs to simplify your automation projects.

This Pump Module is supplied complete with the following
components:
Component Quantity
Main Unit 1
Motor/Encoder Extension Cable, 6 ft. 1
DC Power Extension Cable, 6 ft. 1
RS-232 Cable 1
Symphony Program Diskettes 1
Grease, 6 oz. jar 1


Specifications Back to Top

Specifications702202
# of Syringes1
Accuracy0.5 %
Average Linear Force200 lb (maximum)
Cable Length English6 ft
CommunicationsDC Power 4-pin Header (Friction Lock – Molex or AMP)
RS-232 4-pin RJ-11 Telephone Jack
User I/O 9-pin D-Sub Female
Control Board Mounting HolesMounting holes for (4) #6 screws
Control Board Mounting Length English4.5 in
Control Board Mounting Length Metric11.43 cm
Control Board Mounting Width English7 in
Control Board Mounting Width Metric17.78 cm
Depth English11.875 in
Depth Metric30.2 cm
DisplayOptional 2x20 Line VFD (p/n 2400-235)
Drive: ControlConstant Current (Chopper) Drive, 2A per phase (max.), microstepping (from 1/2 to 1/32)
Drive: Lead Screw Pitch24 threads per inch
Drive: Ratio2:1 (1:2 optional)
Environmental Humidity20 to 80% RH non-condensing
Environmental Operating Temperature Metric0 to +35 °C (natural convection cooling)**Note: operating temperature may be extended with forced air cooling
Environmental Storage Temperatue Metric­20 to +70 °C
Flow Rate Maximum220.82 ml/min (with 140 ml syringe)
Flow Rate Minimum0.001 µl/hr (with 0.5 µl syringe)
Height English6.625 in
Height Metric16.8 cm
KeypadOptional (p/n 2400-252)
Motor1.8° stepper
Mounting HolesMounting holes for (4) #8 screws
Mounting Length English11.375 in
Mounting Length Metric28.9 cm
Mounting Width English5 in
Mounting Width Metric12.7 cm
Net Weight English8.5 lbs
Net Weight Metric3.86 kg
Pump FunctionInfuse/Withdraw
Pusher Travel Rate Maximum190.676 mm/min
Pusher Travel Rate Minimum0.18 µm/min
RS-232RS-232
Reproducibility0.1 %
Step Rate Maximum416.7 µsec/step
Step Rate Minimum27.3 sec/step
Syringe Size Maximum140 ml
Syringe Size Minimum0.5 µl
Voltage Range(+)12 to +40VDC, ±5%, 75W (user supplied)
Width English5.5 in
Width Metric14 cm

Back to Top



Journal Articles Back to Top

Displaying 1 to 10 of 121 results for High Pressure OEM Syringe Pump Modules   Page 1  2  3  4  5  6  7  8  9  10  11  12  13 

Score

Document

Description

100% Login to view PP98 inf 44.pdf
[408 KB]
Calibrated Syringe Pump use in finding Anticonvulsant Properties of Ganazolone
100% Login to view HS08 pressure transducer.pdf
[46 KB]
Pressure Transducer use in finding blood pressure and heart rate spectral changes
89% Login to view PP107 Predictability Modulates Human Brain Response to Reward.pdf
[300 KB]
Dual Syringe Pump use in finding Predictability Modulates Human Brain Response to Reward
76% Login to view PP94 Syringe Pump.pdf
[717 KB]
Aquaporin Deletion in Mice Reduces Intraocular Pressure and Aqueous Fluid Production
75% Login to view PP107 mri pmp5.pdf
[295 KB]
Dual Syringe Pump use in finding Predictability Modulates Human Brain Response to Reward
65% Login to view PP44 Syringe MS Pump 11.pdf
[519 KB]
Comprehensive Two-Dimensional High-Performance Liquid Chromatography for the Isolation of Overexpressed Proteins and Proteome Mapping
58% Login to view PP06 Molecular Clocks.pdf
[108 KB]
PHD 22/2000 Syringe Pump used in finding Molecular Clocks
56% Login to view PP02 Intake of High-Fat Food is Selectively Enhanced by Mu Opioid Receptor Stimulation Within the Nucleus Accumbens.pdf
[215 KB]
Microdrive Pump use in finding Intake of High-Fat Food is Selectively Enhanced by Mu Opioid Receptor Stimulation within the Nucleus Accumbens
56% Login to view PP114 pump microchannel.pdf
[74 KB]
Syringe Infusion Pump use in finding Microchannel Devices for Efficient Contacting of Liquids in Solvent Extraction
55% Login to view PP04 pump ms.pdf
[53 KB]
Analysis of Jojoba Oil with syringe pump 'apparatus 22'
   

Back to Top



Manuals Back to Top

Displaying 1 to 5 of 5 results for High Pressure OEM Syringe Pump Modules  

Score

Document

Description

100% Login to view 702202 OEM Pump Manual.pdf
[602 KB]
High Force Modulat Pumping Component User's Manual
83% Login to view 702217 Nanomite Syringe Pump Manual.pdf
[893 KB]
Nanomite Syringe Pump User's Manual
73% Login to view Nanoleader Pump Series NLS110P Pressure Pump Manual.pdf
[2.2 MB]
705000 NLS110P Pump Manual
72% Login to view 703005 PHD ULTRA Manual.pdf
[2.4 MB]
PHD Ultra Syringe Pumps Manual - Standard/Remote/Programmable
50% Login to view Syringe Selector.pdf
[127 KB]
Manual & Automatic Syringe Selector Manual
   

Back to Top




  Privacy Policy   |   Terms and Conditions   |   Site Map
Copyright©2004 Harvard Apparatus - All rights reserved.
1-800-272-2775   1-508-893-8999
Website-related comments or concerns can be emailed to our webmaster.

Anesthesia Equipment Animal Handling Products Behavior Research Products Biological Measurement Products Cardiovascular Products Cell Biology Products CPK® Atomic Models Delivery & Sampling Products Dialysis / Sample Prep / SPE Electrophysiology Products Electroporation / Transfection General Laboratory Products Homeothermic/Temperature Control Isolated Organ & Isolated Tissue Microincubation / Microinjectors Molecular Biology Products Neuroscience Products Non-Invasive Blood Pressure Physiology Products Respiratory Products Stereotaxic Instruments Surgical Equipment Surgical Instruments Syringe Pumps/Peristaltic Pumps Tubing & Connectors Ventilators Application Specific Product Overviews