Warner Cell Electrophysiology & Cell Biology Research
Hugo Sachs Elektronic Isolated Organ and Tissue Systems
BTX Electroporation and Electrofusion Products
Panlab Behavior Research Products
Hoefer Electrophoresis Products
Clark Glass Products
Navicyte Ussing/Diffusion Chambers
Harvard Apparatus

GoDaddy
 

Home Microincubation / Microinjectors Chamber System Heater Controllers
 


Larger Views

Chamber System Heater Controllers
These heater controllers have been designed to provide quiet, electric power to a variety of perfusion heating devices including Warner Series 20 heater platforms, solution in-line heaters and the DH-35 culture dish heater, see page M92.
We now offer two high temperature modifcation models of this Heater Controller. These model are useful for applications involving the study of the capsaicin receptors.

Search for Product in current catalog
Item Listing Full Description Specifications Journal Articles Manuals
Item# Description In Stock Lead Time U.S. List Price Quantity
640100 Model TC-324B Heater Controller, Single Channel -- 5 days View Pricing
Login to Shop
640101 Model TC-344B Heater Controller, Dual Channel Yes -- View Pricing
Login to Shop
641453 Model TC-324BHT Perfusion Heater Controller Single Channel with High Temp Modification, 65°C -- 5 days View Pricing
Login to Shop
641454 Model TC-344BHT Perfusion Heater Controller Dual Channel with High Temp Modification, 65°C -- 5 days View Pricing
Login to Shop

For local pricing, contact your nearest distributor.



Full Description Back to Top

• Slow ramped DC power for quiet operation
• Temperature set with single control
• Manually controlled DC output
• High Temperature models available

These heater controllers have been designed to provide quiet, electric power to a variety of perfusion heating devices including Warner Series 20 heater platforms, solution in-line heaters and the DH-35 culture dish heater, see page M92.
We now offer two high temperature modifcation models of this Heater Controller. These model are useful for applications involving the study of the capsaicin receptors.
Each channel can supply up to 18 watts into a load of 8 ?. Maintenance of heat setpoint is controlled automatically in AUTO mode via thermistor feedback. A loop-speed selector allows for selection of three feedback speeds to optimize stability for thermal response characteristics of the device being heated. Heat setpoint maintenance may also be manually controlled in MANUAL mode.

Ease of Use
In AUTO mode, the desired temperature is set with a single SET TEMPERATURE control. The connected chamber platform, solution heater or other device is automatically controlled to the set temperature. Accuracy is typically better than ± 1°C and under ideal conditions will approach ± 0.1°C. In MANUAL mode, the controller provides a variable DC output adjustable from zero to +12 volts.

Quiet Operation in Recording Setups
Highly filtered DC supplies and slow-ramped analog switching circuitry deliver power without adding noise to the system.
Thermistor Readouts

Each channel reads two thermistors simultaneously: T1 for the control of the system, and a T2 for any point of interest. These temperatures are read on the meter and also available at front panel outputs for recording devices. Unical thermistors are used and can be replaced without the need for recalibration.


Specifications Back to Top

Specifications640100640101641453641454
CertificationsCECECECE
Depth Metric25.4 cm25.4 cm25.4 cm25.4 cm
Height Metric8.9 cm8.9 cm8.9 cm8.9 cm
Inputs: Rear Panel BNCRear Panel BNC for T2 (TA-29 Thermistor Cable assembly)Rear Panel BNC for T2 (TA-29 Thermistor Cable assembly)Rear Panel BNC for T2 (TA-29 Thermistor Cable assembly)Rear Panel BNC for T2 (TA-29 Thermistor Cable assembly)
Manual Voltage Range0 to 12 V0 to 12 V0 to 12 V0 to 12 V
Max. Ouput Voltage12 V DC12 V DC12 V DC12 V DC
Max. Output Current1.5 Amps1.5 Amps1.5 Amps1.5 Amps
Max. Output Power18 W (8 ? load)18 W (8 ? load)18 W (8 ? load)18 W (8 ? load)
Meter3 1/2 digit LED display of:
Set Temperature, 50°C max.
T1 (Control Thermistor) temp.
T2 (Monitor Thermistor) temp.
Heater Voltage, 12 V max
3 1/2 digit LED display of:
Set Temperature, 50°C max.
T1 (Control Thermistor) temp.
T2 (Monitor Thermistor) temp.
Heater Voltage, 12 V max
3 1/2 digit LED display of:
Set Temperature, 50°C max.
T1 (Control Thermistor) temp.
T2 (Monitor Thermistor) temp.
Heater Voltage, 12 V max
3 1/2 digit LED display of:
Set Temperature, 50°C max.
T1 (Control Thermistor) temp.
T2 (Monitor Thermistor) temp.
Heater Voltage, 12 V max
ModelTC-324BTC-344BTC-324BHTTC-344BHT
Recoder Outputs: Control Temp. OutT1 (Control Thermistor) 100 mV/°CT1 (Control Thermistor) 100 mV/°CT1 (Control Thermistor) 100 mV/°CT1 (Control Thermistor) 100 mV/°C
Recoder Outputs: Monitor Temp. OutT2 (Monitor Thermistor) 100 mV/°CT2 (Monitor Thermistor) 100 mV/°CT2 (Monitor Thermistor) 100 mV/°CT2 (Monitor Thermistor) 100 mV/°C
Weight Metric3.6 kg6.8 kg3.6 kg6.8 kg
Width Metric20.0 cm43.0 cm20.0 cm43.0 cm

Back to Top



Journal Articles Back to Top

Displaying 1 to 10 of 33 results for Chamber System Heater Controllers   Page 1  2  3  4 

Score

Document

Description

100% Login to view Interleukin-12-mediated induction of systemic immunity.pdf
[213 KB]
Interleukin-12-mediated induction of systemic immunity in the perihpery and recruitment of activated T cells into the brain produce limited antitumor effects compared with interleukin-2
77% Login to view PP33PR 1.PDF
[393 KB]
Micro/nano spacecraft thermal control using a MEMS-based pumped liquid cooling system with use of Programmable syringe pump
75% Login to view dispo eq dial3.pdf
[868 KB]
An mRNA structure that controls gene expression by binding S-adenosylmethionine
50% Login to view PP41 y2-Melanocyte-Stimulating Hormone Suppression of Systemic Inflammatory Responses to Endotoxin.pdf
[495 KB]
Melanocyte-stimulating hormone suppression of systemic inflammatory responses
28% Login to view OT01 Chambers.pdf
[963 KB]
Primary Uroepithelial Cultures
28% Login to view Primary Uroepithelial Cultures.pdf
[971 KB]
Primary Uroepithelial Cultures
27% Login to view PP86.pdf
[180 KB]
Dexenfluramine-Induced Systemic Vasoconstriction
25% Login to view PP35 Peristaltic Pump.pdf
[381 KB]
Peristaltic pump use in finding Improvements to parallel plate flow chambers to reduce reagent and cellular requirements
25% Login to view WR16 RC25.pdf
[275 KB]
Recording Chamber use in finding Hypoxic Augmentation of Fast-Inactivating and Persistent Sodium Currents in Rat Caudal Hypothalamic Neurons
24% Login to view HS02IS Evaluation of Skeletonized LITA.pdf
[228 KB]
Isolated Organ Set use in Intraoperative and Laboratory Evaluation of Skeletonized Versus Pedicled Internal Thoracic Artery
   

Back to Top



Manuals Back to Top

Displaying 1 to 3 of 3 results for Chamber System Heater Controllers  

Score

Document

Description

100% Login to view TC 344B Manual.pdf
[499 KB]
Dual Channel Heater Controller User's Manual
60% Login to view Ussing Chamber Systems.pdf
[1.5 MB]
Guide to Ussing Chamber Systems
52% Login to view TC 324B Manual.pdf
[508 KB]
Single Channel Heater Controller User's Manual
   

Back to Top



  Privacy Policy   |   Terms and Conditions   |   Site Map
Copyright©2004 Harvard Apparatus - All rights reserved.
1-800-272-2775   1-508-893-8999
Website-related comments or concerns can be emailed to our webmaster.

Anesthesia Equipment Animal Handling Products Behavior Research Products Biological Measurement Products Cardiovascular Products Cell Biology Products CPK® Atomic Models Delivery & Sampling Products Dialysis / Sample Prep / SPE Electrophysiology Products Electroporation / Transfection General Laboratory Products Homeothermic/Temperature Control Isolated Organ & Isolated Tissue Microincubation / Microinjectors Molecular Biology Products Neuroscience Products Non-Invasive Blood Pressure Physiology Products Respiratory Products Stereotaxic Instruments Surgical Equipment Surgical Instruments Syringe Pumps/Peristaltic Pumps Tubing & Connectors Ventilators Application Specific Product Overviews