T-Type Ultra Fast Thermocouple Probe

T-Type Ultra Fast Thermocouple Probe

The MLT1402 T-type Ultra Fast Thermocouple Probe is a tissue implantable microprobe with an ultra fast response time of 0.005 seconds. It is inserted into tissue with a supplied 23 ga. hypodermic needle. The MLT1402 has a 2-pin T-type thermocouple plug, suitable for connection to the ML312 T-type Pod, ML313 Cardiac Output Pod or ML290 Thermometer for T-type Thermocouples.

Item# Description U.S. List Price Quantity
No options of this product are available.
Login for U.S. Pricing

The MLT1402 T-type Ultra Fast Thermocouple Probe is a tissue implantable microprobe with an ultra fast response time of 0.005 seconds. It is inserted into tissue with a supplied 23 ga. hypodermic needle. The MLT1402 has a 2-pin T-type thermocouple plug, suitable for connection to the ML312 T-type Pod, ML313 Cardiac Output Pod or ML290 Thermometer for T-type Thermocouples.


Operation:

A thermocouple is a temperature measurement sensor that consists of two dissimilar metals joined together at one end (a junction) which produces a small thermoelectric voltage when heated. The MLT1402 uses a type T thermocouple element with a copper/constantan junction. To use the thermocouple, insert the tip of the sensor into a needle from the hub end until the tip is flush with the bevel. Insert the needle to the required depth then withdraw the needle along the lead, pressing gently forward on the sensor to keep it in position. Continue drawing the needle back along the lead until it reaches the plug end of the probe. Tape to secure the thermocouple.

Application:

The MLT1402 T-type Ultra Fast Thermocouple is suitable for temperature measurements to a maximum of 150 °C.