Syringe Heater (SW-10/6)

Syringe Heater (SW-10/6)

Solution reservoir heating is an important technique used to eliminate outgassing of perfusate solutions.

  • Compatible with TC-324C and TC-344C heater controllers
  • Quiet operation
  • Compact design accommodates six 10 cc syringes
  • Designed for situations where a large heated water bath is inconvenient (such as a shared facility)

Item# Description U.S. List Price Quantity
64-0111 Model SW-10/6 Six 10 cc Syringe Heater
64-0165 Model SL-6 Stopcock with Luer Connector, pkg. of 6
64-0161 Model DN/60M Constant Flow Syringe, 60 cc, pkg. of 6
64-0220 Model FR-50 Flow Valve
64-0210 MP-6 Perfusion Manifold, 6 Inputs
64-2400 TC-324C Heater Controller, Single Channel
64-2401 TC-344C Heater Controller, Dual Channel
Login for U.S. Pricing

Solution reservoir heating is an important technique used to eliminate outgassing of solutions prior to their entering a heated perfusion chamber. Since the gas load of a solution has dependence on pressure and temperature, preheating a solution at the final pressure before delivery to the chamber will minimize the occurrence of bubbles in the chamber's bath, even if the solution is allowed to cool en route.

Solution reservoir heaters from Warner Instruments are designed for applications where the use of a large heated water bath is inconvenient, such as a shared facility. Designed to accommodate six 10 cc syringes, the SW-10/6 Heater housing is made of anodized aluminum which is both corrosion resistant and serves as an excellent conductor. These heaters may be mounted on any 3/8 or 1/2 diameter laboratory rod and each unit is supplied with a cable assembly for connection to Warner's TC-324 or TC-344 heater controllers (purchased separately).

Also supplied is a TA-29 thermistor for monitoring the temperature within any syringe and six 10 cc syringes. Solutions will reach set temperature approximately 15 minutes after application of power. Additional useful accessories (purchased separately) include solution manifolds, stopcocks with Luer connectors, ring stands, and perfusion flow valves.

Syringe Size 10 cc
Heater Resistance 10Ω
Voltage Requirement Variable to 12 V max.
Temperature Range ambient to 65°C (limited by temperature controller)
Temperature Accuracy ± 1° C
Physical Dimensions
Body, H x W x D 7.5 x 17.8 x 2.5 cm
Weight 680 g
Cable Length 1.9 m
Warranty 1 year
SW10_6 Manual.pdfSyringe Heater (SW-10/6) Manual