Sterile Disposable Pressure Transducer (DTXPlus)

Sterile Disposable Pressure Transducer (DTXPlus)

Minimize distortion in the pressure waveform and maximize accuracy

Item# Description U.S. List Price Quantity
72-4293 DTXPlus Sterile Disposable Single-Use Transducer, pkg. of 10, with HAI Interface Cable
72-4297 DTXPlus Sterile Disposable Single-Use Transducer, pkg. of 10, with Hugo-Sachs PLUGSYS Interface Cable
72-4294 DTXPlus Sterile Disposable Single-Use Transducer, pkg. of 10, with Grass Amplifiers Interface Cable
72-4295 DTXPlus Sterile Disposable Single-Use Transducer, pkg. of 10, with PROPAQ Interface Cable
72-4296 DTXPlus Sterile Disposable Single-Use Transducer, pkg. of 10, with GOULD 600 Interface Cable
72-4292 Replacement DTXPlus Disposable Single-Use BP Transducer, pkg. of 10 (Interface Cable not included)
73-4479 Manual Pressure Calibrator, Range 0-300 mmHg
73-0096 Three-Way Stopcock Type 956R Stainless Steel
73-0097 One-Way Stopcock Type 9500 Stainless Steel
75-0030 One-Way Stopcock Sterile, FLL to MLL, pkg. of 100
75-0039 Three-Way Stopcock Sterile, pkg. of 100
88-0368 Interface Cable for use with Sterile Pressure Sensors (88-0360); Connects to ADInstruments Bridge Amp
72-2883 Interface Cable, 12 ft (3.7 m), Universal Connector to Tinned Leads (no connector), For Use with Other Equipment, Mating Connector Required
Login for U.S. Pricing

DTXPlus blood pressure transducers minimize distortion in the pressure waveform and maximize accuracy.

  • Individually packaged, sterile, disposable blood pressure transducers
  • Five-year shelf life
  • Stopcocks must be purchased separately
  • Class III according to European Medical Devices Directive 93/42/EEC
Transducer Item # Interface Cable
72-4293 HAI
72-4297 PUGSYS TAM
72-4294 Grass Amplifiers
72-4295 PROPAQ
72-4296 Gould
Specifications
Pressure Range -30 to +300 mmHg
Over Pressure Tolerance -700 to +7,800 mm
Frequency Response >1,200 Hz, 15 bandwidth shall be >200 Hz
Transducer Excitation 1 to 10 V RMS
Transducer Excitation
Input Impedance 

1,100 to 3,450 Ω
Transducer Signal
Output Impedance
270 to 330 Ω
Nominal Sensitivity (gain) 5 µV/V/mmHg
Unbalance (Zero Balance) ±40 mmHg
Accuracy 2% of the reading or ±1 mmHg, whichever is
greater over the operating range
Operating Temperature 15° to 40°C