Pressure Transducer (NL108T4)

Pressure Transducer (NL108T4)

Suitable for both arterial and venous blood pressure monitoring allowing continuous pressure monitoring.

  • Suitable for both arterial and venous blood pressure monitoring
  • The NL108T4 can be attached via a lead (NL108L4) to the NL108A pressure amplifier
  • Enables continuous pressure monitoring
  • Highly rugged and reusable transducer

Item# Description U.S. List Price Quantity
72-4222 NL108T4 Pressure Transducer, pkg of 1
72-4223 Replacement Dome for NL108T4 Pressure Transducer - 72-4222, pkg of 1
72-4224 Replacement Dome for NL108T4 Pressure Transducer - 72-4222, pkg of 10
Login for U.S. Pricing

Suitable for both arterial and venous blood pressure monitoring, the NL108T4 can be attached via a lead (NL108L4) to the NL108A pressure amplifier, allowing continuous pressure monitoring. The NL108T4 is a highly accurate and rugged re-useable transducer, which is used in conjunction with disposable domes (NL108D4). These disposable domes are easy to fill and attach to the transducer body and are available in packs of 10 or individually (one dome is included with each transducer). The components of the transducer are electrically isolated from the transducer housing, which is in turn isolated from the saline solution by the silicone membrane of the dome. The transducer is gold plated, making it easier to clean. It is designed primarily for use during blood pressure monitoring, but operates equally well as a monitor for intracranial, gastrologic or intrauterine pressures, as well as urodynamic measurements.

Specifications
Pressure range 20 to 300 mm Hg
Max. Overpressure 10000 mmHg
Sensitivity 50 µV/V/cm Hg
Resonance Frequency 300 Hz typical (transducer and dome)
Maximum Electrical Excitation 15 V DC or AC
Input Resistance 700 O (or equipment specified)
Output Resistance 1000 O (or equipment specified)
Non-Linearity and Hysteresis max. 0.5% of full scale
Zero Balance max. ±30 mmHg
Thermal Sensitivity 0.15%/ºC
Thermal Zero Shift max. 0.25 mmHg/ºC
Operating / Storage
Temperature Range
+10 to +50ºC / -20 to +70ºC
Insulation Resistance min. 10 O
Leakage Current max. 1.5 µA at 250 V, 50 Hz.
High Voltage Resistance 10000 O between saline in dome and transducer leads
Weight 26 g (without cable)
Length of Cable 0.3 m