Perfusion Pressure in Hepatic Artery

Perfusion Pressure in Hepatic Artery

Perfusion Pressure in Hepatic Artery

Item# Description U.S. List Price Quantity
73-3862 APT300 Pressure Transducer for PLUGSYS Module
73-0500 Lab Stand with Triangular Base Plate (one block clamp included)
73-0065 Transducer Amplifier Module (TAM-A)
Login for U.S. Pricing
Perfusion Pressure in Hepatic Artery
Specifications73-0065
Analog IndicationTAM-A LED bar graph 20 LEDs (+13/-7) for visual check on the signal sensitivity approx. 1V/LED TAM-D 3 1/2-digit LED display
Bridge BalanceThrough 10-turn trimmer coarse adjustment and electronic autozero by push button (LED for error if autozero is not possible)
Bridge Supply Voltage+5V / 50 mA max.
Electrical CalibrationSelectable by switch on front panel:
a) 0 V output signal with switch in position '0'
b) Positive or negative calibration output voltage adjustable with 10-turn trimmer if switch is in position 'CAL'
GainSelectable ranges by internal Jumper: 0.2 to 10, 0.4 to 20, 1 to 50, 2 to 100, 4 to 200, 10 to 500, 20 to 1000, 100 to 5000, 200 to 10000. Fine Adjustment Through 10-Turn Trimmer
Output Low-Pass Filtera) Selectable by switch on front panel for pulsatile output signal: 1, 100, 300 Hz
b) Selectable by internal jumper for mean output signal: 0.1, 0.3 Hz
PLUGSYS Width*2 slot units
Power Supply+5V through connector from PLUGSYS bus system
Signal Outputa) On front panel through BNC socket
Transducer Input6-pin socket with screw lock (binder, Amphenol Tuchel) Differential input circuit, imput impedance 1010?