MicroCell (MC-100)

MicroCell (MC-100)

Designed for spot determination of the oxygen content of small samples of water, blood etc., or as an in-line flowcell.

  • For pO2 measurement of blood or other fluids, or for micro flowcell use
  • Small sample volume of approximately 70 µL
  • Flow through or spot measurement of oxygen concentration
  • Exceptional visibility of sample contents

Item# Description U.S. List Price Quantity
69-3016 MicroCell Model MC 100
69-3025 (SK-100) Replacement glass for MC-100
69-3006 Microcathode Oxygen Electrode Model 1302 with polypropylene jackets
69-3001 Model 782 Dual channel oxygen meter with automatic respirometry software
69-3002 (SI-929) Model SI-7929 Six channel oxygen meter with standard respirometry software
Login for U.S. Pricing

This microcell was designed for use with the 1302 Oxygen Sensor either for spot determination of the oxygen content of small samples of water, blood etc., or as an in-line flowcell. The MC 100 all-glass sample chamber has a capacity of approximately 70 µl.

Samples are injected into the chamber through a 1 mm stainless steel tube fitted with a fine PVC needle-locating collar. Alternatively, the cell can be used as a micro flowcell by connecting the outflow from a respiring preparation. The sample leaves the chamber through a glass lure fitting to which a small bore flexible drainage tube can be connected.

The sample chamber is completely surrounded by a glass water jacket supplied from a constant temperature bath. This provides good temperature regulation while providing excellent visibility of the contents of the chamber. The water jacket screws into an acetal base section and can be easily removed for cleaning.

The 1302 Oxygen Sensor fits into the central electrode holder and the membrane at its tip forms the floor of the sample chamber. Can be used with either the 782 Dual-Channel Oxygen Meter or the 929 Six-Channel Oxygen Meter.

Specifications
Sample Chamber 70 µl
Overall Height 106 mm
Jacket and Cell Glass
Base Section Black acetal