Lab Animal Evacuation System

Lab Animal Evacuation System


The Lab Animal Evacuation System was designed to be used with the multistation in facilities that do not have a built-in evacuation system or a fume hood available. The waste gas from the system may be vented to a 2-inch outside line or pulled through activated charcoal so that the air can be recirculated in to the room.

Item# Description U.S. List Price Quantity
72-3066 Lab Animal Evacuation System
110 VAC, 60 Hz
72-3046 Lab Animal Evacuation System
230 VAC, 50Hz
72-3067 Charcoal Refill
72-6000 Supply Line Adapter (to adapt connector to 3/16- to 1/4-in tubing)
72-8209 Slilcone Adapter (to adapt connector to 3/16- to 1/4-in tubing)
Login for U.S. Pricing

The Lab Animal Evacuation System was designed to be used with the multistation in facilities that do not have a built-in evacuation system or a fume hood available. The waste gas from the system may be vented to a 2-inch outside line or pulled through activated charcoal so that the air can be recirculated in to the room. When using the filtration system, the activated charcoal must be changed on a regular basis, to maintain it’s effectiveness. Typically, 8 to 10 hours of anesthesia delivery can be performed for each new supply of charcoal. The lab evacuation system, will accommodate waste gas supply from 1 to 6 stations.

NOTE: When using in-house suction, a scavenger interface valve must be used when operating more than one station, to reduce the pull of air.

Specifications72-3066
Power110 VAC, 60 Hz