IPL-4 Option VCMR-4 73-4298

IPL-4 Option VCMR-4   73-4298

NEGATIVE PRESSURE VENTILATION CONTROLER WITH PRESSURE REGULATOR TO CORE SYSTEM IPL-4 (OPTION VCMR-4)

Item# Description U.S. List Price Quantity
73-1755 VCM4-R PLUGSYS VENTILATION CONTROL MODULE TYPE 681/1
73-1750 Timer Counter Module (TCM)
Login for U.S. Pricing

Negative Pressure Ventilation Controllerwith Regulator to Core IPL-4

Requires pressurized gas supply from tank or house air inthe range of 2-8 bar (29 – 116 PSI)

For both Positive- and Negative-pressure Ventilation

 

Includes:

  • VCM4-R PLUGSYS Ventilation Control Module for Rabbi tLung Respiration
  • TCM PLUGSYS Timer Counter Module for introduction of periodic Sigh Breaths (hyperinflation)
  • Utilizes 6 Slot Units
Specifications73-1750
Count ModeDeep inspiration is produced after predetermined number of respiratory cycles of VCM. Respiration cycles are counted down on presetting counter inside TCM;
Count Mode Continuedwhen zero is reached, VCM is triggered to perform deep inspiration cycle and counter automatically resets to preset value
Input and Output Trigger SignalsConnections made via jumpers to trigger lines of 96-pin VG connector from internal PLUGSYS system bus
PLUGSYS Width*1 slot unit
Power Supply5 V, 350 mA through connector from PLUGSYS bus system
PresetTime or event presetting is selected on two digit thumbwheel switch (00-99) mounted on front panel; time or count range can be altered with switch SEC/MIN or x1/x10 to select appropriate factor
Time ModeDeep inspiration is produced at predetermined time intervals. Quartz-controlled timer inside TCM runs down from set time to ‘00’ and then resets counter back to selected value;
Time Mode Continuedwhen ‘00’ is reached, VCM is triggered to perform deep inspiration cycle
Warning FunctionTCM incorporates warning function which announces next triggering operation; announcement activated when internal timer or counter counts last 10 seconds or respiratory cycles; function is indicated visually on LED and audibly by built-in loudspeaker.
Warning Function ContinuedPurpose of facility is to remind user that deep inspiration is coming shortly. This can be inhibited by pressing ‘SKIP NEXT’ key, if desired to avoid interference with any measurement during experiment