LUNG WEIGHT MEASUREMENT TO CORE SYSTEM IPL-4 (OPTION LW-4) 73-4299

Item# Description U.S. List Price Quantity
73-3468 HSE-WEIGHT MEASUREMENT SYSTEM LS 60 FOR THE ISOLATED LUNG SIZE 4 (EDEMA BALANCE)
73-1602 Carrier Frequency Bridge Amplifier Module (CFBA)
Login for U.S. Pricing

LUNG WEIGHT MEASUREMENT TO CORE SYSTEM IPL-4 (OPTION LW-4)

Includes:

  • HSE Weight Measurement System
  • CFBA PLUGSYS Carrier Frequency Bridge Amplifier Module
  • Utilizes 2 Slot Units
Specifications73-1602
Analog IndicationThrough LED bar graph + 20 LEDs / - 9 LEDs for visual check on the measurement; sensitivity can be selected between 500 mV / LED and 50 mV / LED
Band Width0 to 1 kHz
Bridge Supply VoltageAC excitation voltage (5 kHz sine wave) internal adj. from 3 - 5 V
ConnectorDIN 41612, 96-pin VG
Electrical Calibration 1CAL 1 adjustable through 10-turn trimmer, negative and positive calibration voltage; calibration signal is superimposed on measurement signal by pressing CAL 1 key
Electrical Calibration 2CAL 2 selected internally through link plugs
1. Positive calibration voltage
2. Negative calibration voltage
3. External calibration voltage from PLUGSYS measuring system
Gain0 to 100, 0 to 1000 or 0 to 10,000, depending on various internal and external gain settings
Gain: ExternalSwitch selectable x1 / x10 and continuously adjustable through 10-turn trimmer from 0 to 100%
Gain: InternalSwitch selectable x10 / x100
Input Voltage Range140 mVrms (internal gain setting x10), 14 mVrms (internal gain setting x100)
Output Low-Pass Filter11 positions 0.01 Hz to 1 kHz selected by rotary switch
PLUGSYS Width*2 slot units
Power Supply+5V, 1.1 A through main connector from PLUGSYS bus system
Signal Input6-pin socket with screw lock (binder, Amphenol-Tuchel)
Signal Output1. Front panel BNC socket
Signal Output MUXAfter pressing MUX (MULTIPLEX) key filtered signal output of amplifier is switched for example to input of digital voltage meter module for
Signal Output MUX 2accurate zero adjustment and calibration setting; function is self-maintained and module previously selected is switched off
Suppression SUPAdjustable with 10-turn trimmer, negative and positive values; suppression can be switched on and off with switch
SynchronizationSwitch selectable master or slave (internal)