IPL-2 Option VCMR-2 73-4280

IPL-2 Option VCMR-2  73-4280

Pressure Ventilation Controller with Regulator to Core IPL-2

Item# Description U.S. List Price Quantity
73-2795 Ventilation Control Module (VCM-R); uses Pressure Regulated Gas source (tank or house air) in the range of 2-8 bars
73-1750 Timer Counter Module (TCM)
73-3448 ALTERNATIVE GAS SUPPLY TO CORE SYSTEM IPL-2 WITH OPTION VCMP (OPTION PFGS-2), 73-3448
Login for U.S. Pricing

Pressure Ventilation Controller with Regulator to Core IPL-2


  • Requires pressurized gas supply from tank or house air in the range of 2-8 bar (29 – 116 PSI)
  • Choose this ventilation option when:
  • performing standard positive- and negative-pressure ventilation studies
  • delivering alternative gases during both negative- and positive-pressure ventilation

Includes:

  • VCM-R PLUGSYS Ventilation Control Module with Regulator
  • TCM PLUGSYS Timer Counter Module for introduction of periodic Sigh Breaths (hyperinflation)
  • Adapter for Alternate Gas Supply f/ IPL-2
  • Utilizes 6 Slot Units
Specifications73-279573-1750
Count Mode‘INSP. CYCLE’: 10, 20, 30, 40, 50, 60. 70. 80 (%)Deep inspiration is produced after predetermined number of respiratory cycles of VCM. Respiration cycles are counted down on presetting counter inside TCM;
Count Mode Continued‘RATE’: 30, 40, 50, 60, 70, 80, 90, 100 (min-1)when zero is reached, VCM is triggered to perform deep inspiration cycle and counter automatically resets to preset value
Input and Output Trigger Signals0 to 4 I/min (no back pressure)Connections made via jumpers to trigger lines of 96-pin VG connector from internal PLUGSYS system bus
PLUGSYS Width*DIN 41612, 96-pin VG1 slot unit
Power Supply100 to 5 mmHg electronically controlled5 V, 350 mA through connector from PLUGSYS bus system
Preset5 slot unitsTime or event presetting is selected on two digit thumbwheel switch (00-99) mounted on front panel; time or count range can be altered with switch SEC/MIN or x1/x10 to select appropriate factor
Time Mode5 V 0.4 A, 24 V 0.4 A through connector from PLUGSYS bus systemDeep inspiration is produced at predetermined time intervals. Quartz-controlled timer inside TCM runs down from set time to ‘00’ and then resets counter back to selected value;
Time Mode ContinuedDEEP INSP. OUT (output TTL level open collector)
EXP./INSP. OUT (output TTL level open collector)
DEEP INSP. IN (input TTL load resistance 2.2 kOhm)
when ‘00’ is reached, VCM is triggered to perform deep inspiration cycle
Warning Function[inspiration = low level] / [expiration = high level] TTL level (open collector)TCM incorporates warning function which announces next triggering operation; announcement activated when internal timer or counter counts last 10 seconds or respiratory cycles; function is indicated visually on LED and audibly by built-in loudspeaker.