IH-5 CORE SYSTEM, BASIC 115 V 73-4397

IH-5 CORE SYSTEM, BASIC 115 V 73-4397

IH-5 CORE SYSTEM, BASIC 115 V 73-4397

Item# Description U.S. List Price Quantity
00-1413 BASIC UNIT FOR ISOLATED HEART SIZE 5 TYPE 843/I 00-1413
73-2802 THERMOSTATIC CIRCULATOR E 103, BATH VOLUME 3 L, 115 V/60 Hz
73-0115 MCP-SB2/115V ROLLER PUMP, 2 CHANNELS, EACH 2-1100 ML/MIN
73-0045 PLUGSYS Case, Type 603
73-3862 APT300 Pressure Transducer for PLUGSYS Module
73-1793 Transducer Amplifier Module (TAM-D)
73-3871 HOLDER FOR APT300 FOR LVP MEASUREMENT FOR IH5
73-0065 Transducer Amplifier Module (TAM-A)
Login for U.S. Pricing

IH-5 CORE SYSTEM, BASIC 115 V 73-4397

If your chosen options require more than 20 Slot Units you will need to choose the case expansion option(Option C-EX).

For a working unit, the core system requires the addition of a selection of one of the perfusion pressure controller systems (SCP or STARLING-IH5), one of the options IH5-RGP or IH5-RAB, and a heart chamber(IH5-CARDIAC or IH5-EPHYS). The working heart-ready configuration also requires the species-appropriate aortic/direct flow measurement option.

Specifications73-004573-386273-179373-0065
Analog IndicationLine Cable, Spare Fuses, Screw driver for fitting Function Modules, Document Folder for Operating Instructions of Function Modules, 3 x BNC-BNC Output linesCETAM-A LED bar graph 20 LEDs (+13/-7) for visual check on the signal sensitivity approx. 1V/LED TAM-D 3 1/2-digit LED displayTAM-A LED bar graph 20 LEDs (+13/-7) for visual check on the signal sensitivity approx. 1V/LED TAM-D 3 1/2-digit LED display
Bridge Balance5.2 x 19 x 17.1 in3 mThrough 10-turn trimmer coarse adjustment and electronic autozero by push button (LED for error if autozero is not possible)Through 10-turn trimmer coarse adjustment and electronic autozero by push button (LED for error if autozero is not possible)
Bridge Supply Voltage132 x 483 x 435 mm2 — 15 V DC (or AC up to 5 kHz)+5V / 50 mA max.+5V / 50 mA max.
Electrical CalibrationMetallic housing, gray plastic coated, Universal 19 inch case with pivoting feet, prepared for rack mounting>1 kHzSelectable by switch on front panel:
a) 0 V output signal with switch in position '0'
b) Positive or negative calibration output voltage adjustable with 10-turn trimmer if switch is in position 'CAL'
Selectable by switch on front panel:
a) 0 V output signal with switch in position '0'
b) Positive or negative calibration output voltage adjustable with 10-turn trimmer if switch is in position 'CAL'
Gain20 slot units max.>5,000 VSelectable ranges by internal Jumper: 0.2 to 10, 0.4 to 20, 1 to 50, 2 to 100, 4 to 200, 10 to 500, 20 to 1000, 100 to 5000, 200 to 10000. Fine Adjustment Through 10-Turn TrimmerSelectable ranges by internal Jumper: 0.2 to 10, 0.4 to 20, 1 to 50, 2 to 100, 4 to 200, 10 to 500, 20 to 1000, 100 to 5000, 200 to 10000. Fine Adjustment Through 10-Turn Trimmer
Output Low-Pass Filter100 to 240 VAC, 50/60 Hz, built-in line filter, 100 W max. consumption–300 to 300 mmHga) Selectable by switch on front panel for pulsatile output signal: 1, 100, 300 Hz
b) Selectable by internal jumper for mean output signal: 0.1, 0.3 Hz
a) Selectable by switch on front panel for pulsatile output signal: 1, 100, 300 Hz
b) Selectable by internal jumper for mean output signal: 0.1, 0.3 Hz
PLUGSYS Width*3-pin IEC mains with fuse and mains switch on back of housing15 to 40 °C2 slot units2 slot units
Power Supply14.3 lbs356 ± 1% Ω+5V through connector from PLUGSYS bus system+5V through connector from PLUGSYS bus system
Signal Output6.5 kg4,000 mmHga) On front panel through BNC socket a) On front panel through BNC socket
Transducer Input± 300 mmHg6-pin socket with screw lock (binder, Amphenol Tuchel) Differential input circuit, imput impedance 1010?6-pin socket with screw lock (binder, Amphenol Tuchel) Differential input circuit, imput impedance 1010?
5 µV/V/mmHg
–25 to 70 °C
PLUGSYS
< 1% /°C
< 0.04 mmˆ3/100 mmHg
< 0.2 mmHg/°C
<25 mmHg
<25 mmHg

Features and Benefits

 

  • Compact design, optimized individually for: rabbit, rat or guinea pig
  • Constant pressure or constant flow perfusion in one unit, easily switch between the two modes
  • Compact, doesn’t require wall installed high water columns, suitable for hearts from hypertensive animals (perfusion pressures up to 300 mmHg are possible)
  • Unique integrated small volume aortic block with built-in bubble trap and compliance chamber
  • Unique cannulae design

- Cannula resistance is optimized according to Hagen-Poiseuilles Physical Law
- Different cannula sizes available, all metal, no fragile glass cannulae

  • Proprietary mini holders allow easy and stress-free access to hold electrodes and catheters in position
  • Drug injection pathway built directly into aortic perfusate stream
  • Easily upgraded to a Working Heart System

 

PARAMETERS ON AN IH-5 BASIC SYSTEM:
The following signals are recorded as raw data:

  • Isovolumetric Left Ventricular Pressure (Balloon Method)
  • Aortic (Perfusion) Pressure
  • Coronary Flow*


The following parameters are calculated from the raw data (using a DAQ System):

  • dLVP/dt, dLVP/dt Max, dLVP/dt Min, Contractility Index
  • Systolic and Diastolic LVP
  • Heart Rate
  • Mean Perfusion Pressure
  • Mean Perfusion Flow*
  • Coronary Resistance*

* This Parameter is based on Indirect Flow Measurement with Option SCP or Direct Flow Measurement with IH5-STARLING and IH5-DIRECT added to the core system.

Features and Benefits

 

  • Compact design, optimized individually for: rabbit, rat or guinea pig
  • Constant pressure or constant flow perfusion in one unit, easily switch between the two modes
  • Compact, doesn’t require wall installed high water columns, suitable for hearts from hypertensive animals (perfusion pressures up to 300 mmHg are possible)
  • Unique integrated small volume aortic block with built-in bubble trap and compliance chamber
  • Unique cannulae design

- Cannula resistance is optimized according to Hagen-Poiseuilles Physical Law
- Different cannula sizes available, all metal, no fragile glass cannulae

  • Proprietary mini holders allow easy and stress-free access to hold electrodes and catheters in position
  • Drug injection pathway built directly into aortic perfusate stream
  • Easily upgraded to a Working Heart System

 

PARAMETERS ON AN IH-5 BASIC SYSTEM:
The following signals are recorded as raw data:

  • Isovolumetric Left Ventricular Pressure (Balloon Method)
  • Aortic (Perfusion) Pressure
  • Coronary Flow*


The following parameters are calculated from the raw data (using a DAQ System):

  • dLVP/dt, dLVP/dt Max, dLVP/dt Min, Contractility Index
  • Systolic and Diastolic LVP
  • Heart Rate
  • Mean Perfusion Pressure
  • Mean Perfusion Flow*
  • Coronary Resistance*

* This Parameter is based on Indirect Flow Measurement with Option SCP or Direct Flow Measurement with IH5-STARLING and IH5-DIRECT added to the core system.