Large Animal Ventilator (Model 613) - Single Animal, Volume Controlled

Large Animal Ventilator (Model 613) - Single Animal, Volume Controlled

This is the workhorse of the Harvard Apparatus Volume Controlled Ventilators. There are literally thousands of these Ventilators in use in leading research laboratories throughout the world. Their design has been essentially unchanged over the years. There are dozens of examples of these Ventilators giving continuous service for more than thirty years with proper care and maintenance.

• For animals from 1.5 to 50 kg (3.3 to 110 lb) in weight
• Volume adjustable from 30 to 750 cc/stroke
• Variable Inspiratory: Expiratory Ratio

Item# Description U.S. List Price Quantity
55-0715 Harvard Large Animal Ventilator, Model 613, 115 VAC
55-0723 Harvard Large Animal Ventilator, Model 613, 230 VAC
55-0731 Overhaul Kit for Harvard Apparatus Large Animal Ventilator; Contains O-rings, Valve Springs, Lubricants, etc. to Overhaul Ventilator* *Overhaul Kit does not include replacement cylinder
Login for U.S. Pricing
This is the workhorse of the Harvard Apparatus Volume Controlled Ventilators. There are literally thousands of these Ventilators in use in leading research laboratories throughout the world. Their design has been essentially unchanged over the years. There are dozens of examples of these Ventilators giving continuous service for more than thirty years with proper care and maintenance.

• For animals from 1.5 to 50 kg (3.3 to 110 lb) in weight
• Volume adjustable from 30 to 750 cc/stroke
• Variable Inspiratory: Expiratory Ratio

Note that the volume adjustment is made on the piston shaft, and that therefore accurate volume adjustment cannot be made while the ventilator is running. However, the rate and the I:E % can be varied while the ventilator is running. Air/gas valving is by the four port system with separate ports for gas inlet, gas outlet (to animal), gas return (from animal), and ventilator exhaust.

The gas inlet can be used to supply air or any non-flammable gas supply. Like the other Harvard mechanical ventilators, attachment of the ventilator exhaust to a column of water can be used to produce a positive and expiratory pressure (PEEP).

Specifications55-071555-0723
CertificationsCECE
Control ModesVolumeVolume
Depth English8.98.9
Depth Metric22.522.5
Displaynonenone
Gas SupplyRoom air or non-flammable mixed gasRoom air or non-flammable mixed gas
Height English11.811.8
Height Metric3030
I:E Ratio25 to 50 of single stroke25 to 50 of single stroke
Model613613
Net Weight English4545
Net Weight Metric2121
Number of Animals11
PEEPProvided via attachment of water columnProvided via attachment of water column
Port Size English OD0.6250.625
Port Size ID English0.50.5
Port Size ID Metric12.712.7
Port Size OD Metric15.915.9
Respiration Rate Maximum5050
Respiration Rate Minimum77
Respiratory Rate NoteAdjustable from 7 to 50 strokes/min while the ventilator is runningAdjustable from 7 to 50 strokes/min while the ventilator is running
Sigh Frequencynonenone
Sigh PressureNot availableNot available
SpeciesRats, Guinea Pigs, Rabbits, Cats, DogsRats, Guinea Pigs, Rabbits, Cats, Dogs
Stroke Volume NoteAdjustable from 30 to 750 ml/stroke (adjustment must be made while ventilator is off)Adjustable from 30 to 750 ml/stroke (adjustment must be made while ventilator is off)
Tital Volume Maximum750750
Tital Volume Minimum3030
Voltage Range115 VAC230 VAC
Weight Range Maximum5050
Weight Range Minimum1.51.5
Width English19.719.7
Width Metric5050
DOCUMENT DESCRIPTION
550715_Ventilator_613_Manual.pdfLarge Animal Volume Controlled Ventilator Manual