V8401 Hand-Held Capnocheck

V8401 Hand-Held Capnocheck

The V8401 Capnocheck is a highly transportable monitor designed to provide spot-check reading of End Tidal CO2 (ETCO2). It also provides inspired C02 and respiration measurement on most animals. Simple one-touch controls make it easy to use. The V8401 operates on internal rechargeable batteries or can be connected to an AC adapter. An infrared link sends data to an external printer.

• Hand-held Capnograph for small animal veterinaryfacilities (e.g., cats)
• Highly portable unit that measures capnography,respiration rate and ETCO2
• Side-stream technology for use on intubated and non-intubated subjects
• One-touch function keys for easy operation

Item# Description U.S. List Price Quantity
72-6017 Hand Held Capnocheck Model V8401, 220 VAC, 50 Hz
72-6016 Hand Held Capnocheck Model V8401, 120 VAC, 60 Hz
Login for U.S. Pricing
The V8401 Capnocheck is a highly transportable monitor designed to provide spot-check reading of End Tidal CO2 (ETCO2). It also provides inspired C02 and respiration measurement on most animals. Simple one-touch controls make it easy to use. The V8401 operates on internal rechargeable batteries or can be connected to an AC adapter. An infrared link sends data to an external printer.

• Hand-held Capnograph for small animal veterinaryfacilities (e.g., cats)
• Highly portable unit that measures capnography,respiration rate and ETCO2
• Side-stream technology for use on intubated and non-intubated subjects
• One-touch function keys for easy operation

The accuracy and reliability of capnometry is available in a lightweight transportable unit that runs on AC or battery power. A bright LED display provide ETCO2 and respiration rate measurements. One-touch, direct function keys and the lightweight ergonomic design offer ease of operation with both intubated and non-intubated animals. The V8401 draws a sample of gas through the sample cell. It uses an infrared (IR) light source, optical band pass filter and detector to measure the amount of CO2 in the sample. The ETCO2 is displayed as an average of 4 breaths. The continuous CO2 waveform is used to detect each breath and to compute the respiratory rate. When using N2O in an anesthesia setup, the N2O compensation can be ‘Enabled’ to compensate for the interference of N2O on the CO2 measurement. The V8401 Capnocheck comes complete with an instruction manual, AC adapter/charger, and protective boot.

Specifications
Display
    LCD, with electroluminescent back light

Indicators
    High, medium and low priority alarm
    Alarm mute
    AC power

Graphic Trends
    ETCO2 and respiration rate

Waveforms
    CO2 waveforms

Numerical Data and Messages
    ETCO2, respiration rate and status messages

Sidestream Capnography
    Measurement: Non-dispersive IR absorption
    Measurement Range: 0 to 10% CO2, STPD
    Units: User-selectable - mmHg, Vol. %, kPa
    Display Range: 0 to 100 mmHg, 0 to 13.3 kPa, 0 to 10% CO2
    Accuracy:
       ±2 mmHg or 4% (<50 breaths/min) whichever is greater
       51 to 80 breaths/min: +4/-10% of reading
       81 to 150 breaths/min: +4/-20% of reading
    Averaging: 4 BPM

N2O Compensation
    Selectable 40% (default = off)

System Response Time
    2.46 sec to 90% with 8 ft patient time

Flow Rate
    120 ±20 ml/min

Calibration
    Manual 2 point

Respiration rate
    0-150 BPM

Averaging
    4 BPM

Serial Output
    Infrared port for printer

Battery
    Lithium Ion, 7.4 VDC
    User Replaceable
    Rechargeable
    Continuous life of 6 hrs
    2.5 hrs charging time

Dimensions, H x W x D
    12.7 x 11.1 x 7.4 cm (5 x 4.38 x 2.9 in)

Weight
    0.63 kg (22 oz)