Disposable BP Transducer

Disposable BP Transducer

The MLT0670 is a low-cost, light-weight, disposable blood pressure transducer that gives consistent and accurate readings. The transducer is supplied in sterile packaging, however MUST be sterilized prior to use with human subjects. The transducer is suitabe for use with the FE117 BP Amp using the MLAC05 Adapter Cable.

Item# Description U.S. List Price Quantity
77-0064 Disposable BP Transducer (Stopcock), Model MLT0670
Login for U.S. Pricing

The MLT0670 is a low-cost, light-weight, disposable blood pressure transducer that gives consistent and accurate readings. The transducer is supplied in sterile packaging, however MUST be sterilized prior to use with human subjects. The transducer is suitabe for use with the FE117 BP Amp using the MLAC05 Adapter Cable.


Operation:

Before use, finger-tighten all connections. Blood pressure is transmitted to the sensor via a subject cannula or catheter. Before connecting to a catheter, prime the system by flushing three or four times until all visible air bubbles are removed. Use a 10 mL syringe if continuous flushing is not available. Open the 3-way stop cock to air to take a zero reading for the transducer before calibration and recording blood pressure.

Application:

The MLT0670 Disposable BP Transducer is designed for measuring arterial and venous blood pressure in humans and large animals.
Specifications77-0064
Cable Length English9.8 in
Cable Length Metric250 mm
Certifications*na
ConnectorSpacelabs plug
Control Port*na
Dimensions H x W x D English*na
Dimensions H x W x D Metric*na
Excitation2 to 10 V DC
Frequency Response*na
Input Accuracy*na
Input Amplification Ranges*na
Input Connection Type*na
Input Impedance350 Ω ± 10%
Input ImpedanceSee product page for complete specifications
Input Isolation Rating*na
Input Isolation*na
Input Leakage Current*na
Input Safety*na
Input Sensitivity*na
Input User Offset*na
Input Zeroing and Offset*na
Length English3.4 in
Length Metric85 mm
Maximum Power Needs*na
ModelMLT0670
Non linearity and Hysteresis±2 % of reading or ±1 mmHg (whichever is greater)
Operating Conditions*na
Operating Life>500 hours
Operating Pressure Range?50 to +300 mmHg
Operating Temperature15 °C to +40 °C
Output Analog Signal*na
Output Impedance<400 Ω
Overpressure protection-400 to +4000 mmHg
Package1
Product FamilyData Acquisition
Sensitivity5 ?V/V/mmHg
Storage Temperature-25 °C to +70 °C
Transducer Alternative Types*na
Transducer Excitation*na
Transducer Type*na
Weight English*na
Weight Metric*na
Weight(including cable): 26 g
Width English1.6 in
Width Metric40 mm