Blood Pressure Transducer/Cable Kit

Blood Pressure Transducer/Cable Kit

The MLT1199 BP Transducer/Cable Kit comprises an MLT0699 Disposable BP Transducer and a MLAC06 DIN(8) to Spacelabs Adapter Cable for connection to a Bridge Amp.

Item# Description U.S. List Price Quantity
77-0124 Blood Pressure Transducer/cable kit, Model MLT1199
Login for U.S. Pricing
The MLT1199 BP Transducer/Cable Kit comprises an MLT0699 Disposable BP Transducer and a MLAC06 DIN(8) to Spacelabs Adapter Cable for connection to a Bridge Amp. The individual components may be ordered separately.

Operation: Before use, finger-tighten all connections. Blood pressure is transmitted to the sensor via a subject cannula or catheter. Before connecting to a catheter, prime the system by flushing three or four times until all visible air bubbles are removed. Use a 10 mL syringe if continuous flushing is not available. Adjust the 3-way tap to take a zero reading for the transducer before recording blood pressure.

Application: The MLT0699 Disposable Blood Pressure Transducer is designed for measuring arterial and venous blood pressure. For animal use only.

Specifications77-0124
Amplification Input AmplifierSee product page for complete specifications
Body Length English2.2 in
Body Length Metric57 mm
Body Width English1.5 in
Body Width Metric39 mm
Cable Length English3.9 ft
Cable Length Metric1.2 m
ConnectorSpacelabs plug
Excitation2 to 10 V DC
Input Impedance350 Ω ±10%
MLAC06 Adaptor Cable8 pin DIN to Spacelabs plug
ModelMLT1199
Non linearity and Hysteresis±2 % of reading or ±1 mmHg (whichever is greater)
Operating Life>500 hours
Operating Range-50 to +300 mmHg
Operating Temperature15° C to 40° C
Output Impedance