Aerosol Nebulizer

Aerosol Nebulizer

The Aerosol Nebulizer is ideal for nebulizing drugs sensitive to ultrasonic cavitation.

  • Low particle sizes (100% when particles are below 10 µm)
  • No solution warming required
  • Aerosol is automatically transported by compressed air
  • Connecting block required to attach Nebulizer to HSE-HA single/double chamber plethysmograph or mouse plethysmograph
  • Multigas Inlet Adapter (73-2919) required for use with MicroVent, MiniVent or MidiVent
  • Available with Multi-Gas Adapter Kit for Chamber/Box use

Item# Description U.S. List Price Quantity
73-1963 Aerosol Nebulizer
73-3433 Aerosol Nebulizer with Multi-Gas Adapter Kit for Chamber/Box Use
73-2919 Multi-Gas Inlet Adapter for MicroVent, MiniVent and MidiVent
73-3300 Tubing Connection Kit for Aerosol Nebulizer
Login for U.S. Pricing

The Aerosol Nebulizer is ideal for nebulizing drugs sensitive to ultrasonic cavitation

  • Low particle sizes (100% when particles are below 10 µm)
  • No solution warming required
  • Aerosol is automatically transported by compressed air
  • Connecting block required to attach Nebulizer to HSE-HA single/double chamber plethysmograph or mouse plethysmograph
  • Multigas Inlet Adapter (73-2919) required for use with MicroVent, MiniVent, or MidiVent
  • Available with Multi-Gas Adapter Kit for Chamber/Box use

This aerosol jet nebulizer requires an operating pressure of approximately 1.5 bar (22 psi) from a compressed air source. All of the particles generated by the jet nebulizer are 10 µm or less in size with 60% of the particles being 2.5 µm or less. A special connecting block is used to attach the nebulizer to the HSE-HA single and double chamber plethysmographs and are configured in a number of respiratory mechanics applications.

Specifications

This aerosol jet nebulizer requires an operating pressure of approximately 1.5 bar (22 psi) from a compressed air source. All of the particles generated by the jet nebulizer are 10 µm or less in size with 60% of the particles being 2.5 µm or less. A special connecting block is used to attach the nebulizer to the HSE-HA single and double chamber plethysmographs and are configured in a number of respiratory mechanics applications.

Size Micron Below Band in Micron % in Band
10.5 100 13.6 to 10.5 0
8.19 99.7 10.5 to 8.19 0.2
6.37 98.0 8.19 to 6.37 1.7
4.97 92.2 6.37 to 4.97 5.8
3.88 80.9 4.97 to 3.88 11.3
3.04 68.3 3.88 to 3.04 12.6
2.40 61.5

3.04 to 2.40

6.8
1.90 57.7 2.40 to 1.90 3.8
1.52 54.5 1.90 to 1.52 3.2
1.22 52.0 1.52 to 1.22 2.5
    1.22 to 0 52.0