Advisor® Vital Signs Monitor

Advisor® Vital Signs Monitor

The Advisor® Vital Signs Monitor is the leading choice in multiparameter monitoring and is ideal for larger animals such as cats, dogs, pigs, horses, and non-human primates. It has a large easy-to-read display, intuitive design, and is available in multiple configurations. Standard parameters include ECG, pulse oximetry, heart rate, non-invasive blood pressure. Optional parameters include end tidal CO2, respiratory rate, 2-channel temperature, and 2-channel invasive blood pressure (IBP).

• Four waveform color display
• Point and click interface
• View all parameter trending information on screen
• 2-year warranty

Item# Description U.S. List Price Quantity
72-6077 Advisor® 3 Lead ECG, SpO2 Heart Rate and NIBP
75-0259 Advisor® 3-Lead ECG, SpO2, Heart Rate, NIBP, IBP, and Temperature (Includes Temperature Probe and Cable, IBP Transducer and Cable)
75-0260 Advisor® 3 Lead ECG, SpO2, Heart Rate, NIBP, End Tidal CO2 and Respiration Rate
75-0261 Advisor® 3 Lead ECG, SpO2, Heart Rate, NIBP, 2-Channel IBP, End Tidal CO2, 2-Channel Temperature and Respiration Rate (Includes Temperature Probe and Cable, Reusable Esophageal/ Rectal Temperature Probe, IBP Transducer and Cable, and Capnograph start-up k
72-3233 CO2 Calibration Kit
72-3246 Moisture Trap
72-3254 Reflectance Sensor (Small)
72-3256 Universal Y Sensor (Small to Medium)
72-3258 C-Clip Pulse Ox Sensor
72-6003 Mini Clip Oximetry Sensor
72-3157 Pulse Oximeter Sensor Cable, 5 ft
75-0264 3-Lead ECG Wires with Alligator Clips, 40 in
72-6079 3-Lead ECG Wires with Alligator Clips, 60 in
72-6080 3-Lead ECG Cable (Can be used with 75-0264, 75-0265)
72-6011 Temperature Cable
59-9808 Small Flexible Vinyl Rectal/Esophageal Thermister Probe, pkg. of 1 (B)
75-0249 Invasive Blood Pressure Transducer Cable
75-0248 Invasive Blood Pressure Transducer
72-3253 CO2 Monitor Exhaust Kit for Scavenging Halogenated Gasses
Login for U.S. Pricing
The Advisor® Vital Signs Monitor is the leading choice in multiparameter monitoring and is ideal for larger animals such as cats, dogs, pigs, horses, and non-human primates. It has a large easy-to-read display, intuitive design, and is available in multiple configurations. Standard parameters include ECG, pulse oximetry, heart rate, non-invasive blood pressure. Optional parameters include end tidal CO2, respiratory rate, 2-channel temperature, and 2-channel invasive blood pressure (IBP).

• Four waveform color display
• Point and click interface
• View all parameter trending information on screen
• 2-year warranty

All Advisor models come complete with power cord, reflectance Y sensor, oximetry extension cable, 3 lead ECG, NIBP cuff set and manual.

Specifications
Displays & Controls:
   Main Display 10.4 in diagonal LCD, 640 x 480 pixels
ECG:
   Heart Rate Range 20-350 BPM
   Heart Rate Accuracy ±2 BPM or ±2% whichever is greater
QRS Detection Range:
   Normal 0.5 to 5 mV
   High Sensitivity 0.1 to 5 mV
   Alarm Ranges Heart    Rate 20-350 BPM and Off
   Lead Selection I, II, or III
   Display Gain Settings X0.5, X1, X2, X4
SpO2:
   SpO2 Range 0-100% functional saturation
   SpO2 Accuracy ±2% @ 70-100%, <70% unspecified
   Pulse Rate Range 20-350 BPM
   Pulse Rate Accuracy ±2 BPM or ±2% whichever is greater
   SpO2 Alarm Ranges 50-100% and Off
NIBP:
   Method of    Measurement Oscillometric with step-down deflation
   Systolic Pressure 40 to 265 mmHg
   Mean Arterial 27-222 mmHg
   Disastolic Pressure 20-200 mmHg
   Pulse Rate Range 25 to 350 BPM
   Pulse Rate Accuracy 25-250 BPM ±2% or 3 BPM whichever is greater
251-300 BPM ±5%, <25 or >300 unspecified
   Inflation Pressure    Settings 200 mmHg Cat, 150 mmHg Dog, 120 mmHg Horse
   Adjustment Range 60-280 mmHg Cat, 120-280 mmHg Dog/Horse
   NIBP Clinical    Accuracy Based on human algorithm and meets accuracy requirements for ANSI/AAMI SP10: 1992 and 2002
   Auto Interval Times 1-10, 15, 20, or 30 minutes; 1, 2, 4 or 8 hrs
   Alarm Range 1 mmHg Steps and Off
   Patient Safety Max cuff inflation time is 75 sec; max blood pressure reading time is 120 sec
Capnography:
   Measurement Non-Dispersive IR Absorption
   Measurement Range 0-10% CO STPD
   Display Range 0 to 100 mmHg; 0 - 14kPa; 0-10% CO2
   Accuracy ±2 mmHg or 4% of reading whichever is greater
   N2O Compensation Selectable 40%
   Averaging 4 breaths
   Flow Rate 150 ml/min ±20 ml
   ETCO2 Alarm Range 0-100 mmHg, 0-14 kPa, 0-10% CO2 and Off
Respiration Rate:
   Range 0-150 BPM
   Accuracy ±1 BPM
   Alarm Range 3-120 BPM and off
2-Channel IBP:
   Measuring Range -50-300 mmHg
   Accuracy ±1 mmHg or ±1%, whichever is greater
   Alarm Limits -50-300 mmHg
Pulse Rate:
   Range 20-350 BPM
   Accuracy ±BPM or ±2% whichever is greater
   Averaging 8 sec
2-Channel Temperature:
   Range 0° - 50°C
   Accuracy ±0.1°C plus temperature sensor tolerance
   Resolution ±0.1°C
   Alarm Limits 20° to 50°C and Off
   Sensor Compatibility YSI 400 Series Temperature Sensors
Power Requirements:
   AC Input 100-240 VAC, 0.8 A, 50/60 Hz
   Battery Operating    Time 60 -120 minutes
Dimensions
   (H x W x L)
30.5 x 18.0 x 34.0 cm (12.0 x 7 x 13.25 in)
Weight 4.1 - 5 kg (9-11 lb)